Explanation of Graphical Symbols :
Sterilization of Fiber-Optic Probe :
After each treatment, the fiber-optic probe should be disinfected or sterilized. Wrap
completely in an autoclave sterilization pouch, then autoclave at 132°C/75 psi for 4
minutes by using a pre-vacuum sterilizer to get maximum results. Allow the fiber-optic
probe to air-dry for 20-30 minutes in the sterilization pouch before removing.
IMPORTANT :
Disconnect the fiber-optic probe from the handpiece and autoclave by itself.
Product Cleaning:
If there is any dirt or stains on the surface of control panel, handpiece or other
components, please use a cloth to wipe clean. If it is hard to clean, please use
a neutral cleansing agent. 70%
isopropyl alcohol is recommended.
IMPOR
TANT :
Never u
se a solvent or volatile oil for cleaning; this can lead to fading
of the pr
int on the casing or the metal ring falling off from the probe.
Class II equipment
ISO 7000-1641 : Follow operating instructions or consult instructions
for use
Cooling System :
>> This device is designed with a built-in cooling fan which automatically activates
whenever the device is in use, and continues to run for a few seconds after the
cycle has ended.
>> If the temperature gets exceeds
45
°C, the fan will continue to run until the
internal temperature has cooled down.
>> If the temperature gets exceeds
65
°C, the device will automatically shut off and
the fan will continue to run. Until the internal temperature has cooled down, the
device will be able to restart.
>> If the unit needs to be used for continuous curing, please decrease the curing time.
>> Under the premise of not touching the oral tissue, when using the curing light or
switching the treatment area or switching any operation settings, it is necessary
to cool down for seconds before starting the next curing cycle in case the curing
light being used for over 15 seconds during a curing cycle for more cycles.
IEC 60601-1
:
Applied part classification - Complies with standards
protecting against electric shock for type B equipment
IEC 60417-5031 : Direct current.
IEC 60417-5009 : STAND-BY
IMPORTANT :
Every component of this device is reusable, but only the fiber-optic probe
needs to be disinfected or sterilized. Other components can not be disinfected
or sterilized.
ULTIMATE BASE280
• Sicherheitsmassnahmen.................................................................................. 39
• Erläuterung der grafischen Symbole................................................................ 40
• Komponenten und Funktionen......................................................................... 42
• Installation........................................................................................................ 43
• Verwendung Des.............................................................................................. 44
• Signale Des ......................................................................................................46
• Problembehebung.............................................................................................47
• Spezifikationen:.................................................................................................48
Inhalt:
Disclaimer:
• Vorgesehener Einsatzzweck............................................................................ 39
1.
Vielen Dank, dass Sie sich für den Kauf der Polymerisationslampe BA
Ultimate BASE280 LED entschieden haben. Bitte lesen Sie sich die
Gebrauchsanweisung sorgfältig durch und beachten Sie bei der Bedienung
des Geräts die darin enthaltenen Vorschriften. Bitte bewahren Sie die
Gebrauchsanweisung zur späteren Verwendung sorgfältig auf.
2.
BA International übernimmt keine Verantwortung für
Fehlfunktionen oder Schäden, die durch die unsachgemäße Bedienung,
Entfernung von Teilen, Modifizierung oder inkorrekte Wartung des Geräts
entstehen.
3.
Wir übernehmen keine Verantwortung für Fehlfunktionen oder Schäden, die
durch die Verwendung des Geräts in einer ungeeigneten Umgebung
entstehen.
4.
Wir können nicht für Fehlfunktionen oder Schäden haftbar
gemacht werden, die durch die unsachgemäße Bedienung, inkorrekte
Wartung oder höhere Gewalt entstehen.
5.
BA International behält sich das Recht vor, jegliche Verantwortung für
Produktdefekte abzulehnen, die durch Naturkatastrophen, abnormale
Temperaturen oder Luftfeuchtigkeit verursacht wurden, die nicht den hierin
enthaltenen Vorgaben entsprechen.
38