5 Operation
Operating Instructions BA ULTIMATE
20
65 56 083 D3560
.
201.10.02
.
02
5
Operation
Use only undamaged burs
NOTICE! Use only burs and diamond polishers that are sharp and
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
Drill Fixation _ Turbine
CAUTION! A loose or partially removed bur can detach itself from the
head or break off. This may cause injury! Therefore, use the turbine only
if the bur is inserted at least 10 mm and is clamped securely in place.
Overheating at the preparation site
CAUTION! Insufficient cooling leads to overheating of the
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
Pulling back the patient's cheek _ Turbine
CAUTION! Do not pull the patient's cheek back with the handpiece
while the turbine is rotating. This would actuate the pushbutton, thus
creating a risk of burning the patient's oral mucosa.
Observing manufacturer documentation
IMPORTANT: For application details and operating data of the different
preparation tools, please refer to the information provided by the
manufacturer.
6556083.book Seite 20 Montag, 18. April 2016 4:35 16
Summary of Contents for BA 695
Page 2: ...6556083_Titel fm Seite 2 Montag 18 April 2016 5 16 17 ...
Page 4: ...6556083 book Seite 2 Montag 18 April 2016 4 35 16 ...
Page 40: ...6556083 book Seite 38 Montag 18 April 2016 4 35 16 ...
Page 42: ...6556083 book Seite 2 Montag 18 April 2016 4 30 16 ...
Page 78: ...6556083 book Seite 38 Montag 18 April 2016 4 30 16 ...
Page 80: ...6556083_FR book Seite 2 Montag 18 April 2016 4 40 16 ...
Page 116: ...6556083_FR book Seite 38 Montag 18 April 2016 4 40 16 ...
Page 118: ...6556083_ES book Seite 2 Montag 18 April 2016 4 51 16 ...
Page 156: ...6556083_IT book Seite 2 Montag 18 April 2016 5 02 17 ...
Page 192: ...6556083_IT book Seite 38 Montag 18 April 2016 5 02 17 ...
Page 194: ...6556083_NL book Seite 2 Montag 18 April 2016 5 12 17 ...
Page 230: ...6556083_NL book Seite 38 Montag 18 April 2016 5 12 17 ...