Operating Instructions BA ULTIMATE
7 Conditioning
65 56 083 D3560
.
201.10.02.02
35
ohne DAC PROFESSIONAL_Turbine
Approved steam sterilizers are those that fulfill the requirements of
either EN 13060, class B or EN 13060, class S and which are also
suitable for the sterilization of highspeed handpieces.
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
After sterilizing - heading
After sterilizing
Follow-up _ Turbine
1. Remove the turbine from the steam sterilizer immediately.
CAUTION! The turbine will be hot. This may cause burns!
NOTICE! Do not immerse the turbine in cold water to speed up the
cooling process. This will damage your turbine!
2. Store all turbines so that they are protected from contamination.
3. Sterilize again once the storage period has elapsed.
6556083.book Seite 35 Montag, 18. April 2016 4:35 16
Summary of Contents for BA 695
Page 2: ...6556083_Titel fm Seite 2 Montag 18 April 2016 5 16 17 ...
Page 4: ...6556083 book Seite 2 Montag 18 April 2016 4 35 16 ...
Page 40: ...6556083 book Seite 38 Montag 18 April 2016 4 35 16 ...
Page 42: ...6556083 book Seite 2 Montag 18 April 2016 4 30 16 ...
Page 78: ...6556083 book Seite 38 Montag 18 April 2016 4 30 16 ...
Page 80: ...6556083_FR book Seite 2 Montag 18 April 2016 4 40 16 ...
Page 116: ...6556083_FR book Seite 38 Montag 18 April 2016 4 40 16 ...
Page 118: ...6556083_ES book Seite 2 Montag 18 April 2016 4 51 16 ...
Page 156: ...6556083_IT book Seite 2 Montag 18 April 2016 5 02 17 ...
Page 192: ...6556083_IT book Seite 38 Montag 18 April 2016 5 02 17 ...
Page 194: ...6556083_NL book Seite 2 Montag 18 April 2016 5 12 17 ...
Page 230: ...6556083_NL book Seite 38 Montag 18 April 2016 5 12 17 ...