7
EN
DE FR ES IT NL
CAUTION
•
Hold the spray can upright.
•
Be sure to hold the handpiece firmly to prevent slipping
out of hand by the spray pressure when lubricating.
•
Supply lubricant until it comes out of the handpiece
head.
•
If you wish to expel excessive oil from inside of the
handpiece, rotate the handpiece for approx.15 seconds
without a Rubber Cup / Brush.
WARNING
• If blood infiltrates inside a handpiece, an
automatic handpiece cleaning and lubrication
system may not totally clean the internal
handpiece components, which may lead to
internal blood coagulation. Coagulated blood
inside the handpiece may cause handpiece
failure and overheating, causing burn
injuries. To avoid the risk, B.A. International
recommends to immediately lubricate those
handpieces with BA ULTIMATE SPRAY.
01_BA65E_EN.indd 7
2019/10/28 18:23:50
Summary of Contents for BA OPTIMA BA65E
Page 2: ...ノド 1 2 3 4 BA65E_図出力 indd 1 2019 10 28 18 30 08 ...
Page 3: ...BA65E_図出力 indd 2 2019 10 28 18 30 08 ...
Page 15: ...12 01_BA65E_EN indd 12 2019 10 28 18 23 51 ...
Page 39: ...36 03_BA65E_FR indd 36 2019 10 28 18 25 37 ...
Page 51: ...48 04_BA65E_ES indd 48 2019 10 28 18 26 32 ...
Page 63: ...60 05_BA65E_IT indd 60 2019 10 28 18 27 42 ...