Guidance and manufacturer's declaration – electromagnetic immunity
The model
BASE800
is intended for use in the electromagnetic environment specified below. The
customer or the user of the model
BASE800
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for
input/output
lines
±
2 kV for power
supply lines
±
1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s) to
earth
±
1 kV line(s) to line(s)
±
2 kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
interruptions and
voltage
variations
on power supply
input lines
IEC 61000-4-11
<5 %
(>95% dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95% dip in
U
T)
for 5 sec
U
T
<5 %
(>95% dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95% dip in
U
T)
for 5 sec
U
T
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
model
BASE800
requires continued
operation during power
mains
interruptions, it is recommended
that the model
BASE800
be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.