Operating Instructions BA ULTIMATE
6 Follow-up
65 56 067 D3560
.
201.09.03.02
31
6
Follow-up
6.1
After each treatment session
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
✔ The bur instrument must not be moving.
✔ Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur instrument with tweezers.
3. Pre-disinfect directly at the treatment center [ → 33].
4. Remove the turbine.
5. Transport the turbine to the hygiene room in a suitable transport
container.
6556067_EN.book Seite 31 Montag, 27. November 2017 12:27 12