5 Operation
Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
26
64 01 553 D3551
.
201.03.03
.
02
5
Operation
Use only undamaged burrs
NOTICE! Use only burrs and diamond polishers that are sharp and
undamaged. Use clean burrs and diamond polishers to avoid dirt in the
clamping system.
Clamping the burr _ instrument
CAUTION! A loose or partially removed burr instrument can come
loose from the head or break off. Risk of injury! Therefore the instrument
must only be used when the burr instrument is securely clamped at least
10 mm deep.
Only operate when the clamping system is closed
CAUTION! The handpiece must only be operated with the clamping
system closed.
Overheating at the preparation site
CAUTION! Insufficient cooling leads to overheating of the
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
6401553.book Seite 26 Montag, 13. Juni 2016 4:27 16