7 Conditioning
Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
42
64 01 553 D3551
.
201.03.03
.
02
1. Brush the instrument under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and
adapters according to manufacturer's instructions, e.g. with WL
Clean made by ALPRO
®
.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds. You can use
e.g. WL-dry made by ALPRO
®
.
5. Conduct thermal disinfection or unwrapped steam sterilization.
6. Maintain mechanical parts manually [ → 43].
7. Maintain the push button chuck manually [ → 45].