7 Conditioning
Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
46
64 01 553 D3551
.
201.03.03
.
02
Procedure
Vorgehensweise _ Winkelstück
✔ The instrument is clean and disinfected.
1. Press the contra-angle handpiece head with the chuck firmly against
the spray can nozzle.
2. Spray the chuck for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
3. Wipe any spray that comes out with a disinfection cloth.
ULTIMATE SPRAY
BA
6401553.book Seite 46 Montag, 13. Juni 2016 4:27 16