5
5) Wipe the handpiece dry.
*If the dirt could not be removed from the hole, clean it by brush.
7-2 Cleaning (Handpiece)
1) Remove dirt and debris from the product. Do not use a wire brush.
2) Wipe clean with alcohol-immersed cotton swab or cloth.
This icon denotes that the product can be washed via Thermo
Disinfector.
Refer to the Thermo-Disinfector manual.
•
After washing with Thermo-Disinfector and prior to lubrication,
dry the product until all internal moisture is thoroughly removed.
Thermo-Disinfector moisture remaining inside the product could
reduce the effect of lubrication and could cause corrosion inside
of the product.
•
To clean the product never use any solvent such as benzine or
thinner.
CAUTION
7-3 Lubrication
BA ULTIMATE SPRAY
Apply BA ULTIMATE SPRAY every time after each use and/or before
autoclaving.
1) Remove the handpiece from the hose.
2) Remove the bur from the handpiece.
3) Mount the Tip Nozzle into the spray can port. (Fig.7)
4) Insert the Tip Nozzle into the Drive Air Port of the handpiece. Hold
the handpiece and spray for approximately 2-3 seconds. Apply
lubricant until it expels from the handpiece head for at least 2
seconds. (Fig.8)
•
When applying spray be sure to hold the handpiece firmly to
prevent the handpiece from slipping out of the hand due to the
spray pressure.
•
Hold the spray can upright.
CAUTION