10
8. OPERATION PROCESS
Operation Method
1. How to charge
• Plug the Power cord jack into the inlet placed back side of Charging
Station.
• Insert the Power cord and plug in.
• Turn on the Power switch and check if the power lamp is on and
Beep sound is ringing Place the motor handpiece into the Charging
Station. Charging Lamp is on and starts to be charged.
• Charging lamp lights Green when the battery is fully charged and
lights Orange while it charged. The Orange light is blinking when it
has short-circuited or battery needs to be changed
Warning
• Never use the Charging Station for other purpose except the motor
handpiece
• Charging time is normally 2.5 hours, but it depends on battery us-
age conditions, battery freshness, ambient temperature, etc. Older
batteries are especially prone to signifi cantly shorter charging and
operating times.
• Batteries may warm up during charging, but it is not a failure.
• The battery is defective if the Orange light is blinking during
charging. Charge the battery after replacing with a new one.
2. Changing Contra Angle
Contra angle can be connected to Motor handpiece at any position.
Align Contra angle positioning pins with the positioning slots of Mo-
tor handpiece and insert the head until “Click” Pull it out axially for
removing.
Warning
Check if the Contra angle head connected properly to the motor hand-
piece. Otherwise, the motor handpice is not operating properly.
3. Changing File
• Insert the File to the chuck
• Push it into until “Click” sound.
• File removing: Press Push Button and pull the fi le out
Notice
• Charging Lamp may change from Orange to Green light quicker
than normal when charging, depending to the battery charge level.
• To keep the motor handpiece putting into Charging Station for a long
time, it may not be charged properly.
• If the Orange lamp is blinking due to short circuit, contact dealer
or manufacturer.
• Charge the motor handpiece during its inactive time.
Warning
• Turn off the power to remove or connect the contra angle head.
• Pull the fi le slightly after inserting, to check if the fi le has been
inserted properly.
• Clean the shank of the fi le prior to insertion.
Allowing dirt to enter the chuck could cause loss of concentricity and
deterioration of gripping force.
BAE370R_IFU-Final.indd 10
02/04/2019 12:22:32
Summary of Contents for Optima E+
Page 2: ...2 BAE370R_IFU Final indd 2 02 04 2019 12 22 31...
Page 16: ...16 BAE370R_IFU Final indd 16 02 04 2019 12 22 34...
Page 30: ...30 BAE370R_IFU Final indd 30 02 04 2019 12 22 36...
Page 44: ...44 BAE370R_IFU Final indd 44 02 04 2019 12 22 39...
Page 58: ...58 BAE370R_IFU Final indd 58 02 04 2019 12 22 41...
Page 72: ...72 BAE370R_IFU Final indd 72 02 04 2019 12 22 44...
Page 86: ...86 NOTES BAE370R_IFU Final indd 86 02 04 2019 12 22 46...
Page 87: ...87 NOTES BAE370R_IFU Final indd 87 02 04 2019 12 22 46...