English
11
10
1. Only use the
Ultimate 1400
after it has completed an adequate hygiene procedure.
2. Never place the
Ultimate 1400
handpiece, light guide, or depleted battery packs on the recharging
stations before they are cleaned and disinfected.
3. Disinfect with alcohol-based fluids by wiping the complete exterior of the
Ultimate 1400
components.
Do not submerge in fluids and be careful to avoid getting the battery pack contacts wet. Keep battery
contacts dry for proper operation.
4. The battery pack can be removed after the
Ultimate 1400
exterior has been cleaned and disinfected.
5.
Battery packs and handpiece may not be autoclaved!!
6. General Care & Maintenance
• Regularly inspect the
Ultimate 1400
for damage and if necessary send for repair service.
• Use the built-in radiometer to monitor output.
• The surface of the recharging station can be cleaned and disinfected using a mild cleaner or disinfectant with an
alcohol base. To avoid scratches, always use a soft, lint-free cloth to clean plastic surfaces.
6.1. Battery Care
• As is the case with all batteries, battery capacity will decline during the course of its product lifetime. Batteries are a
consumable item. This is due to the internal chemical reaction inside batteries. Never open the batteries. Replace-
ment is at user’s discretion.
• When the
Ultimate 1400
is not in use, battery packs should be kept on the charging mounts to ensure maximum
battery operating time.
• Battery contacts do not require cleaning and must be kept dry.
• This device is provided with 2 batteries. The
Ultimate 1400
and battery should not be disposed of in normal domes-
tic waste. For environmental reasons, please dispose of the device and battery according to local environmental
guidelines or regulations.
• When the Ultimate 1400 is not in use for a prolonged period of time, the battery pack should be disconnected from
the unit and/or connected to the recharger unit to ensure maximum battery operating time.
7. Trouble-Shooting
If there is any doubt about safe operation, then power must be disconnected and the battery removed to avoid ac-
cidental operation. This may be the case, for example, if the apparatus is visibly damaged or no longer works correctly.
8. Compliance
8.1. Declaration of Conformity
BA International hereby declares that the design, structure and version introduced to the dental market of the unit de-
scribed in this manual correspond to the relevant safety and health requirements of the EU Directive. This declaration is
invalid in the event of an alteration made to the unit without prior authorization by BA International.
1. Battery pack contacts do not require cleaning and must be kept dry.
2. Never open or modify the batteries.
3. Keep batteries a safe distance from sources of heat and fire.
Symptom
Action
Unit does not turn “On”
1.) Check battery pack is properly inserted and in the “On” position
2.) Replace battery pack
6.2. Repair Service
A.) Inform your local dental supplier or BA International of a malfunction and the need for repair.
B.) Pack all components of the
Ultimate 1400
in a carton with appropriate protective packaging. Include your name,
address, and a brief description of the problem. Important: enclose a copy of proof of purchase (ex. invoice, ship-
ping note).
C.) Send your package to your local dental supplier or the closest BA International Service Center with a reliable postal/
shipping service.