INSTRUCTION MANUAL
Technical changes reserved
B+B 0141 0315-21 June 2011
B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
21 / 25
Parameter
Function
Title
Name of the process
Input 1 und 2
Informs about the inputs used by channel 1 and 2 respectively
Start
Defines the strategy for the logging start. Options are:
Immediately
The logger starts logging as soon as configuration is sent to the logger
By „data/time“
Start in defined data and time, always after current time. It is possible to perform daily repetitions.
If this option is selected, a new box to define the stop logging time is displayed
Start
Press the „Start-Now“ button from the Monitoring screen to start logging
LogBox
Press the „Start-Now“ button from the Monitoring screen, the Palmtop should be pointed towards the
logger
Digital Input
Readings are performed while the digital input is enabled (closed / 1) and interrupted when the digital
input is disabled (open / 0)
Stop
Defines logging stop mode, options are:
Full memory Loggings can be stored up to the logger full memory capacity is reached
Wrap around Logging never stops. The LogBox will keep on recording the readings and when the
memory is full it will overwrite the oldest record in a circular or wrap around manner
After loggings
The logging will stop after the number of readings here defined
By data/ time
Logging is stopped on user-defined day and time
Interval
Defines the interval between readings: hour, minute and second. For mean, maximum
and minimum values, the shortest interval between loggings is 10 seconds.
Channel 1
Opens the Input 1 Settings screen
Tag
Defines a name for Channel 1
Input
Informs the input type used in Channel 1
Unit
Defines the unit of the variable. For 0-20 mA, 4-20 mA, 0-50 mV and 0-10V the user should write the required unit
Logging mode
It defines how the value measured will be logged. Options are:
Instantaneous
1 reading and 1 logging at each reading Interval.
Average
10 readings at each reading interval. The average value of readings is the value recorded.
Minimum
10 readings at each reading interval. The lowest value found is recorded
Maximum
10 readings at each reading interval. The highest value found is recorded
Lower/Upper
range value
Allows the user to define the reading range for the 0-20 mA, 4-20 mA, 0-50 mV and 0-10V inputs.
Offset
This parameter is used to correct small known mistakes the input signal may
present, such as during sensor switching, transmitter replacement, etc.
Alarms
Enables an alarm that is triggered according to user-defined
parameters
Cancel and OK
Buttons cancel and save configurations defined in Channel 2 screen
Channel 2
Has the same parameters as described for Channel 1
Clocks
Provides access to Logger clocks. When a new
configuration is sent to the logger, clocks are updated
Battery
Defines the moment when the logger turns on the battery switch, before each
reading is performed. Time (up to 10 seconds) can not exceed the mean time
between measurements
Click Apply
Click apply to send the new configuration to the Logger, returning to the Monitoring screen