Wrap around
Readings are continuous, replacing the oldest records with new ones as the number of
readings exceeds the memory capacity.
After
The logger stops readings after a number of readings defined in the corresponding field is
reached.
Date
Readings stop at user-predefined date and time. In case the logger memory capacity is
reached before the date defined, readings are also stopped
Channels
Register Card with parameters related to temperature measurement
Unit
Defines the unit of the value measured: ºC or ºF
Value
Defines how the value measured will be registered. 3 Options are available:
Instantaneous
The logger takes one measurement at the end of the measurement interval defined in
“Period” field and save the resulting value in the internal memory. The minimum interval in
this case is 1 seconds.
Maximum
The logger takes 10 equidistant measurements within the predefined measurement
interval and saves the highest value (Maximum) in the internal memory. In this option the
minimum interval between measurements (Period) is 10 seconds.
Minimum
The logger takes 10 equidistant measurements within the predefined measurement
interval and saves the lowest value (Minimum) in the internal memory. In this option the
minimum interval between measurements (Period) is 10 seconds.
Offset
Gives the user the possibility to automatically correct the values logged.
Alarm
Defines limit values that, when exceeded, characterize an alarm condition. Alarm events
are informed to the user through a flashing mode of the Alarm Indicator LED.
After filling all the fields select OK and settings will be sent to the logger
Note: After clicking the OK button all previous settings and previous data logged are lost.
8.
DATA TRANSFER AND VISUALISATION
Data collected are sent from logger to PC. Data can be collected any time, at the end of the acquisition process or while they are
being acquired. If data collection takes place during the acquisition process, the process will not be interrupted, following the logger
configuration.
It is important to "wake up" the logger before starting communication.
8.4.
Offloading data
Data offload is accomplished by clicking the button
Offload Data
:
During data transfer, a status bar indicates remaining data to be transferred. Data offloading time is proportional to the number of
readings logged.
8.5.
Visualizing data
At the end of values transfer, data are automatically displayed in a graphical format.
8.6.
Chart view
It is possible to select a region of the chart to zoom in. Zoom commands can be accessed through the
View
menu or through zoom
icons from the task bar.
It is also possible to select an area from the chart to zoom in by clicking and dragging the mouse, thus creating a zoom region
starting from the upper left corner on the chart area.
The graphic curves can be vertically dragged with a right-click and dragging the mouse up and down