In this dialog box the user can define the logger operation mode and also obtains general information about the device. The different 
Fields and their meaning are listed in the following: 

 

Title 

In this field, the user identifies the logger by assigning it a name. 

General 

information 

Area with information about the logger, such as Model, Serial Number, Logger Date/Time, 

PC Date/Time, firmware version, memory capacity and number of acquisitions stored in 
memory, etc. In this field, time is constantly updated  while the communication between 
logger and computer is taking place 

Acquisition 

Register  card  where  a  series  of  parameters  that  define  the  acquisition  process  are 
determined. 

Period 

Defines the interval between readings: The minimum interval is one (1) second and the 
maximum is eighteen (18) hours. 
Note: When the type of  value logged is maximum or minimum values (see 

Value

), the 

minimum interval possible is 10 seconds. 

Daily Operations 

Here the user defines the time at which  daily logs will take place 

Estimated time 

This  field  informs  the  user  how  long  it  will  take  to  fill  totally  the  memory,  under  the 

conditions set up during configuration. 

Start of readings 

Readings can be started in one of five different modes 

Immediately  

Measurement starts as soon as set up is ready and sent (OK) to the logger. 

Start via Palm 

Measurement starts with a command sent via a PDA, which runs the software Log Chart 
Palm-OS.  

Date 

Readings start at a predefined date and time. 

Via Start/ Stop 

Measurement

 

starts and ends by pressing the Start/Stop button, in the frontal part of the 

logger, for two seconds. 

Setpoint 

Acquisition  starts  when  a 

temperature

  setpoint  is  reached.  In  this  option,  the  setpoint 

value is defined in the Channels field, where the Alarm parameter is replaced by setpoint.

 

Stop loggings 

Defines  the  event  that  initiates  the  end  of  measurement.  The  following  options  are 
possible. 

Full Memory 

Measurement is stopped when the memory is full (16 000 values).

 

Summary of Contents for TagTemp

Page 1: ...gTemp Temperature Data Logger Technical changes reserved B B Thermo Technik GmbH Heinrich Hertz Str 4 D 78166 Donaueschingen 0141 0315 84 22 06 2016 Fon 49 771 83160 Fax 49 771 831650 info bb sensors com bb sensors com 1 14 ...

Page 2: ...ADVICE 4 4 WASTE AND DISPOSAL ADVICE 4 5 USAGE ACCORDING TO PURPOSE 4 6 PRODUCT DESCRIPTION 5 1 1 Supply voltage 5 6 2 Unpacking 5 6 3 Identification 5 7 OPERATION 6 7 1 Drivers and software installation 6 7 2 Running LogChart_II 7 7 3 Configuring the logger 8 8 DATA TRANSFER AND VISUALISATION 10 8 4 Offloading data 10 8 5 Visualizing data 10 8 6 Chart view 10 8 7 Sheet view 12 8 8 General informa...

Page 3: ... service centers or your salesperson We go after giving you rapid and competent help to minimize the risk of long time outfalls The following operating Instruction are an indispensable part of of this Product It contains important advices for the starting up and further use of the device 2 GENERAL INFORMATION This documentation contains informations which must be paid in attention to assure a high...

Page 4: ... W21 Caution agaist mobile engines W29 Caution against moving parts Electronic waste 3 SECURITY ADVICE The B B Therm Thechnik GmbH assumes no liability for damages occurred through failure to observe these security advices A usage non conforme to the instructions given in this manual can damage the device Body and property damages The use of the measurement system must be restricted to qualified p...

Page 5: ...ata logger TagTemp is a highly autonome temperature measurement device It is supplied with a built in 3 6 V lithium battery CR2032 6 2 Unpacking The following items must be included in the package The electronic logger TagTemp A CD ROM with the Logchart II software and a USB driver Quick logger operation guide The present operating instructions Optional a communication interface upon request 6 3 I...

Page 6: ...30 seconds of user inactivity To reactivate it press again the Start Stop Button The logger operation mode set up is defined in advance by using the LogChart II software Each setting must be defined and the consequences observed The logger starts and stops logging as defined in settings 7 1 Drivers and software installation In the Ir Link3 RS232 model there is a RSR232 Ir communication interface I...

Page 7: ...e same installation wizard will be displayed and you will have to select the option Install software automatically recommended as the driver may already be in the computer Logchart II Software installation Insert the supplied CD ROM in the CD drive of the PCs Laptop The software installlation wizzard will start automatically If it does not please run the LC_II_Setup exe program provided from your ...

Page 8: ...wnload the actual contents of the logger s memory Press to view online the 10 last logged values 7 3 Configuring the logger Make sure the communication interface is connected to the PC port selected The interface must be constantly directed towards the communication window on front part of the logger at a maximum distance of 0 5 m The logger must be first woken up by pressing once the Start Stop S...

Page 9: ... 1 second and the maximum is eighteen 18 hours Note When the type of value logged is maximum or minimum values see Value the minimum interval possible is 10 seconds Daily Operations Here the user defines the time at which daily logs will take place Estimated time This field informs the user how long it will take to fill totally the memory under the conditions set up during configuration Start of r...

Page 10: ...tion the minimum interval between measurements Period is 10 seconds Offset Gives the user the possibility to automatically correct the values logged Alarm Defines limit values that when exceeded characterize an alarm condition Alarm events are informed to the user through a flashing mode of the Alarm Indicator LED After filling all the fields select OK and settings will be sent to the logger Note ...

Page 11: ...Temp Temperature Data Logger Technical changes reserved B B Thermo Technik GmbH Heinrich Hertz Str 4 D 78166 Donaueschingen 0141 0315 84 22 06 2016 Fon 49 771 83160 Fax 49 771 831650 info bb sensors com bb sensors com 11 14 ...

Page 12: ...lar form press the icon Table View Values are presented in a tabular form listing the measurement time and value 8 8 General information sheet This window shows some information about the logger whose data were just read and its configuration The screen can be displayed by pressing the Parameter Viewing icon 8 9 Exporting data Logged data can be exported to different file formats for further analy...

Page 13: ...s correctly selected there are no other programs using the same port during the communication attempts there is no physical obstacle blocking the IR signal the cable is well connected to the PC the port selected does not present any problem 10 MAINTENANCE The housing can be cleaned with a wet cloth Do not use chemical solvents With the exception of exchanging empty batteries the logger is widely m...

Page 14: ...ions 30 x 47 x 0 47 mm Logger PC data transfer time According to the number of logs 40 seconds for 16 000 logs PC Interface Ir USB or Ir Serial LogChart ll software operation environment Set up Software for Windows 95 98 NT 2000 and XP Menus in Portuguese English or Spanish Sets up reads and displays data on the screen TagTemp 0568 0037 Data Logger for the record of up to 16 000 temperature values...

Reviews: