background image

TagTemp USB Temperature Logger 
0568 0037-01 
 

 

0141 0315-92                 28.11.2022 

B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen

 

Seite 

13

 von 

14

 

Technische Änderungen vorbehalten 

Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com

 

 

 

When handling the device and its battery, observe the following 

 

When the device is opened, avoid touching the electronic components and the printed circuit board to 

avoid electrostatic discharge damages. 

 

Pay close attention to battery polarity. 

 

When closing, the cover and its sealing must be mounted correctly and carefully to ensure the 

enclosure sealing. 

 

 

Used batteries must be disposed in accordance to local regulations. Don’t try to recharge, dismount or heat, since 

explosion may occur 

 

10

 

M

OST FREQUENTLY ENCOUNTERED PROBLEMS

 

 

The LED is not flashing: 

The LED flashing light is intentionally weak, and it can be difficult to see it in bright environments. Be sure  it 

is really not flashing. 

 

Communication with the device fails: 

 

o

 

Make sure the COM port is selected correctly and there is no other program using the same port during communication 

attempts. 

o

 

Make sure there is no physical obstacle blocking the infrared signal. 

o

 

Make sure the cable is well connected to the PC port. 

o

 

Make sure the port selected does not present any problem. 

 

11

 

Q

UESTIONS

For further information about this or other products of B+B Thermo-Technik GmbH please do not hesitate to contact us at: 

 

B+B Thermo-Technik GmbH

 

Heinrich-Hertz-Straße 4

 

 

78166 Donaueschingen

 

 

Deutschland

 

 

Tel.: +49 771 83160

 

Fax: +49 771 831650

 

 

E-Mail: info@bb-sensors.com

 

 

www.bb-sensors.com 

 

Summary of Contents for TagTemp

Page 1: ...Operating Manual TagTemp Temperature Data Logger 0568 0037 01 V1 0x G B B Thermo Technik GmbH Heinrich Hertz Str 4 D 78166 Donaueschingen ...

Page 2: ...on 4 7 NXperience Software 4 7 1 Installation 4 7 2 Running NXperience 4 7 2 1 General Parameters 5 7 2 2 Channels Parameters 5 7 2 3 Configuration of LOG Register 6 7 2 4 Finalization 6 7 3 Downloding and Data Visualisation 7 7 3 1 Downloading Data 7 7 3 2 Visualizing Data 7 8 LOGCHART II Software 8 8 1 Installing LogChart II 8 8 2 Running LogChart II 8 8 3 Configuring the Device 9 8 4 Stop Pause...

Page 3: ...which must be paid in attention to assure a highly effective and secure use of the supplied product Please read through the following instructions and make yourself familiar with the handling of the product before you insert it in your processes Keep this document always readily to hand so you can consult it by need 2 1 Labels Symbol Meaning Descripton Advice It is necessary to read the following ...

Page 4: ... the operating temperature will cause damages in the electronic circuits and also damage the housing USAGE ACCORDING TO PURPOSE Please use the product only for the purposes for which it is conceived In case of doubt please first contact B B Thermo Technik GmbH DISOPOSAL Please return the device to B B Thermo Technik GmbH after expiration of its lifetime 3 INTRODUCTION TagTemp USB is a small portab...

Page 5: ... 8 F max over the full scale Note An eventual constant systematic error can be compensated using the OFFSET parameter in the NXperience or LogChart II software Measurement Resolution 0 1 C 32 18 F Memory Capacity 32 000 32 k logs Measurement Interval 5 seconds min to 18 hours max Power Supply 3 0 V lithium battery CR2032 built in Battery Autonomy estimated More than 400 days 1 minute acquisition i...

Page 6: ...ces must be observed The device starts and stops registering as defined in the settings 6 DEVICE CONFIGURATION To perform a configuration the device must be connected to the USB port of the PC 7 NXPERIENCE SOFTWARE 7 1 Installation NXperience software allows you to explore all the features of the device by communicating through its USB interface It is also a complete tool to perform the analysis o...

Page 7: ...device when it was read by NXperience 7 2 2 Channels Parameters TEMPERATURE Tag Allows you to set a name which will be used as channel identification during a download for the channel The field allows up to 16 characters Input Type Shows the type of sensor supported by the device Unit Allows you to define the unit C or F Mode Allows you to define how the value measured will be registered Options a...

Page 8: ...to 3 C 5 4 F can be set for temperature setting This allows to decrease and even zero the temperature measurement error 7 2 3 Configuration of LOG Register LOG Log Interval Allows you to select the frequency in hours minutes or seconds with which a log must be made and logged in the memory Estimated Time Based on the value set in the Log Interval shows the estimated time for the memory to fill MEM...

Page 9: ...omplished by clicking the button Download Logs in the Download screen from NXperience During data transference a status bar indicates remaining data to be transferred Data downloading time is proportional to the number of readings logged 7 3 2 Visualizing Data At the end of values transfer the data will be displayed in a graphical form 7 3 2 1 Chart Window It is possible to select a region of the ...

Page 10: ... some information about the device whose data were just read and its configuration 7 3 2 4 Other Features Check the NXperience Operations Manual for more information on the other features offered by the software such as exporting to other formats filter logs chart junction etc 8 LOGCHART II SOFTWARE 8 1 Installing LogChart II The LogChart II application is used to configure the parameters for data...

Page 11: ...information about the device such as Model Serial Number Data Logger Date Hour Actual Date Hour Firmware Version Memory Capacity and Number of acquisitions stored in memory In this field time is constantly updated while the communication between device and computer is taking place 3 Readings Presents a series of parameters that define the acquisition process a Interval Defines the interval between...

Page 12: ...t values that when exceeded characterize an alarm condition Alarm events are informed to the user through a flashing mode of the Alarm Indicator After filling all the fields select OK and the settings will be sent to the device 8 4 Stop Pause Resume Buttons These buttons will be enabled once the device is fully configured The Stop button allows stop logs permanently so that the device will only re...

Page 13: ... chart to zoom in Zoom commands can be accessed through the View menu or through zoom icons from the task bar It is also possible to select an area from the chart to zoom in by clicking and dragging the mouse thus creating a zoom region starting from the upper left corner on the chart area The graphic curves can be vertically dragged with a right click and dragging the mouse up and down 8 6 2 Data...

Page 14: ... Indicates the level of battery voltage When the battery reaches a low level voltage immediate replacement is recommended to ensure that the device will be continue to function properly Device Status Summarizes the state of the device including its memory and the occurrence of alarm events to the time of the opening of the Diagnostic page 9 BATTERY REPLACEMENT Replace battery with Panasonic lithiu...

Page 15: ...charge dismount or heat since explosion may occur 10 MOST FREQUENTLY ENCOUNTERED PROBLEMS The LED is not flashing The LED flashing light is intentionally weak and it can be difficult to see it in bright environments Be sure it is really not flashing Communication with the device fails o Make sure the COM port is selected correctly and there is no other program using the same port during communicat...

Reviews: