Validated cleaning and disinfection procedure

Mechanical cleaning/disinfection with manual pre-cleaning

Note

The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).

Note

The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.

Manual pre-cleaning with a brush

D–W:

Drinking water

RT:

Room temperature

*Recommended: BBraun Stabimed

Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.

Phase I

Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.

Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.

If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.

Mobilize non-rigid components, such as set screws, links, etc. during cleaning.

Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.

Phase II

Rinse/flush the product thoroughly (all accessible surfaces) under running water.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Mechanical alkaline cleaning and thermal disinfecting

Machine type: single-chamber cleaning/disinfection device without ultrasound

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)

*Recommended: BBraun Helimatic Cleaner alkaline

Check visible surfaces for residues after mechanical cleaning/disinfecting.

Inspection, maintenance and checks

Allow the product to cool down to room temperature.

After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components).

Dry the product if it is wet or damp.

Repeat cleaning and disinfection of products that still show impurities or contamination.

Check that the product functions correctly.

Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service,  see Tech-
nical Service.

Assemble dismountable products, see Assembling.

Check for compatibility with associated products.

Packaging

Appropriately protect products with fine working tips. 

Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected.

Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers).

Ensure that the packaging provides sufficient protection against recontamination of the product during storage.

Steam sterilization

Note

The product can be sterilized either in disassembled or in assembled condition.

Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets).

Validated sterilization process

Steam sterilization using fractional vacuum process

Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665

Sterilization using fractional vacuum process at 134 °C/holding time 5 min

When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity
of the steam sterilizer specified by the manufacturer is not exceeded.

Sterilization for the US market

Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.

Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave.

To achieve a sterility assurance level of 10

-6

, Aesculap recommends the following parameters:

*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an
Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles
may also be suitable, however individuals or hospitals not using the recommended method are advised to validate
any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility
after processing, such as a wrap, pouch, etc.

WARNING for the US market
If this device is/was used in a patient with, or suspected of having Creutzfeldt-Jakob Disease (CJD), the device
cannot be reused and must be destroyed due to the inability to reprocess or sterilize to eliminate the risk of
crosscontamination.

Storage

Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.

Technical Service

For service and repairs, please contact your national B. Braun/Aesculap agency.

Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and for-
feiture of applicable licenses.

Service addresses

Aesculap Technischer Service
Am Aesculap-Platz 
78532 Tuttlingen / Germany
Phone:

+49 (7461) 95-1602

Fax:

+49 (7461) 16-5621

E-Mail:

ats@aesculap.de

Or in the US: 
Attn. Aesculap Technical Services 
615 Lambert Pointe Drive 
Hazelwood 
MO, 63042 
Aesculap Repair Hotline 
Phone:

+1 (800) 214-3392 

Fax:

+1 (314) 895-4420

Other service addresses can be obtained from the address indicated above.

Accessories/Spare parts

For MINOP inVent tubular shaft instruments FH635R, FH636R, FH637R, FH638R, 
FH639R

Validated procedure

Specific requirements

Reference

Manual pre-cleaning 
with brush and subse-
quent mechanical 
alkaline cleaning and 
thermal disinfection

Cleaning brush: e.g., TA012889

20 ml disposable syringe

Place the product in a tray that is suitable 
for cleaning (avoiding rinsing blind 
spots).

Connect components with lumens and 
channels directly to the rinsing port of 
the injector carriage.

Keep working ends open for cleaning.

Place products in the tray with their 
hinges open.

Chapter Mechanical cleaning/disin-
fection with manual pre-cleaning 
and sub-chapter:

Chapter Manual pre-cleaning 
with a brush

Chapter Mechanical alkaline 
cleaning and thermal disinfecting

Phase

Step

T
[°C/°F]

t
[min]

Conc. 
[%]

Water 
quality

Chemical

I

Disinfec-
tant

                      

cleaning

RT (cold)

>15

2

D–W

Aldehyde-free, phenol-free, 
and QUAT-free concentrate, 
pH ~ 9*

II

Rinsing

RT (cold)

1

-

D–W

-

Phase

Step

T
[°C/°F]

t
[min]

Water 
quality

Chemical

I

Prerinse

<25/77

3

D–W

-

II

Cleaning

55/131

10

FD-W

Concentrate, alkaline: 

pH  =  13

<5 % anionic surfactant

0.5 % working solution

pH = 11*

III

Intermediate 
rinse

>10/50

1

FD-W

-

IV

Thermal disin-
fecting

90/194

5

FD-W

-

V

Drying

-

-

-

According to the program for cleaning 
and disinfection device

CAUTION

Damage (metal seizure/friction corrosion) to the product caused by insufficient 
lubrication!

Prior to function checks, lubricate moving parts (e.g. joints, pusher compo-
nents and threaded rods) with maintenance oil suitable for the respective ster-
ilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray 
JG600 or STERILIT® I drip lubricator JG598).

Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*

Sterilization method

Temp.

Time

Minimum drying time

Prevacuum

270 °F/275 °F

4 min

20 min

WARNING

Risk of injury and/or malfunction!

Do not modify the product.

Art. no.

Designation

FH633R

MINOP inVent hand grip for tubuluar shaft instruments

FH635200

MINOP inVent tubuluar shaft instruments outer tube

FF435R

Jaw part with MINOP mirco-scissors Sh/Sh

FF436R

Jaw part with MINOP mirco-scissors Bl/Bl

FF437R

Jaw part with MINOP micro cupped biopsy forceps

FF438R

Jaw part with MINOP micro grasping and preparation forceps

FF439R

Jaw part with MINOP toothed micro forceps

Summary of Contents for Aesculap 012978

Page 1: ...eschrijving Buisschachtinstrumenten Bruksanvisning Teknisk beskrivning R rskaftinstrument N vod k pou it Technick popis N stroje s trubkov m d kem Instrukcja u ytkowania Opis techniczny Rozk adane Ins...

Page 2: ...chnician is responsible for this The recommended chemistry was used for validation Note If there is no final sterilization then a virucidal disinfectant must be used Note For the latest information on...

Page 3: ...ve a sterility assurance level of 10 6 Aesculap recommends the following parameters Aesculap has validated the above sterilization cycle and has the data on file The validation was accomplished in an...

Page 4: ...nts 3773 Corporate Parkway Center Valley PA 18034 USA TA Nr 012978 11 12 V6 nd Nr 45456 Art no Designation FF432R Handle only for instruments FF385R FF389R FF433R Outer tube only for FF385R FF389R FF4...

Page 5: ...erung wurde die empfohlene Chemie verwendet Hinweis Wenn keine abschlie ende Sterilisation erfolgt muss ein viruzides Desinfektionsmittel verwendet werden Hinweis Aktuelle Informationen zur Aufbereitu...

Page 6: ...altezeit 5 min Bei gleichzeitiger Sterilisation von mehreren Produkten in einem Dampfsterilisator Sicherstellen dass die maxi mal zul ssige Beladung des Dampfsterilisators gem Herstellerangaben nicht...

Page 7: ...tie qu apr s validation pr a lable du proc d de traitement st rile La responsabilit en incombe l exploitant au responsable du traitement st rile Pour la validation les produits chimiques recommand s o...

Page 8: ...ateur la vapeur selon DIN EN 285 et valid selon DIN EN ISO 17665 St rilisation par proc d du vide fractionn 134 C dur e de maintien de 5 min En cas de st rilisation simultan e de plusieurs produits da...

Page 9: ...te caso la responsabilidad recaer en el usuario responsable de dicho proceso Para la validaci n se han utilizado las sustancias qu micas recomendadas Nota Deber utilizarse un agente de limpieza viruci...

Page 10: ...en un esterilizador a vapor Asegurarse de que no se sobrepasa la carga m xima del esterilizador a vapor permitida por el fabricante Almacenamiento Almacenar los productos est riles en un envase con b...

Page 11: ...dio medico chirurgico pu essere assicurata soltanto previa validazione nel processo di preparazione La responsabilit di ci ricade sul gestore preparatore Per la validazione stata utilizzata la chimica...

Page 12: ...a vapore accertarsi che non venga superato il carico massimo ammesso per la sterilizzatrice secondo le indicazioni del produttore Conservazione Conservare i prodotti sterili in un imballo ermetico ai...

Page 13: ...sucedido deste produto m dicos ap s a valida o pr via do processo de reprocessamento Nesta situa o o utilizador pessoa encarregue do reprocessamento assume toda a responsabilidade pelo reprocessament...

Page 14: ...egundo a DIN EN ISO 17665 Esteriliza o pelo processo de v cuo fraccionado a 134 C tempo de n o contamina o de 5 min No caso de esteriliza o simult nea de v rios produtos num esterilizador a vapor asse...

Page 15: ...en succesvolle reiniging en desinfectie van dit medische hulpmiddel uitsluitend kan worden gegarandeerdna een voorafgaande validatie vanhet reinigings endesinfectieproces Hiervooris de gebruiker hetre...

Page 16: ...oc d Stoomsterilisator conform DIN EN 285 en gevalideerd conform DIN EN ISO 17665 Sterilisatie volgens gefractioneerd vacu mproc d bij 134 C verblijftijd 5 min Wanneer meerdere producten tegelijk word...

Page 17: ...utf r beredningen har ansvaret f r detta F r valideringen anv ndes den rekommenderade kemikalien Tips Om ingen avslutande sterilisering genomf rs m ste ett desinfektionsmedel med virucid verkan anv nd...

Page 18: ...Se till att maximal till ten last i ngsterilisatorn enligt tillverkarens anvisningar inte verskrids F rvaring F rvara sterila produkter skyddade mot damm i bakteriet t f rpackning i ett torrt m rkt u...

Page 19: ...t bbraun com FF399R FF399R 2 10 9 7 2 7 1 4 3 4 5 2 2 1 2 1 5 5 3 10 9 7 9 7 4 5 5 7 9 8 8 9 7 6 10 5 2 10 9 7 7 1 8 5 5 2 1 9 5 2 5 2 2 4 7 3 10 4 Aesculap Extranet www extranet bbraun com Aesculap 6...

Page 20: ...C 5 B Braun Aesculap Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de TA012889 20 T C F t I 15 2 pH 9 II 1 T C...

Page 21: ...FF387R FF388R FF389R TA Nr 012978 11 12 V6 nd Nr 45456 FH633R MINOP inVent FH635200 MINOP inVent FF435R MINOP FF436R MINOP FF437R MINOP FF438R MINOP FF439R MINOP FF432R FF385R FF389R FF433R FF385R FF3...

Page 22: ...hoz validaci proces pravy Zodpov dnost za to nese provozovatel subjekt prov d j c pravu K validov n byly pou ity doporu en chemik lie Upozorn n Pokud se neuskute n z v re n sterilizace je nutno pou t...

Page 23: ...j v robce Skladov n Steriln v robky skladujte v obalech nepropou t j c ch choroboplodn z rodky chr n n p ed prachem v such m tmav m a rovnom rn temperovan m prostoru Technick servis V ot zk ch servisu...

Page 24: ...zostanie przeprowadzona sterylizacja ko cowa nale y skorzysta z rodka wirusob jczego Notyfikacja Aktualne informacje odno nie przygotowania i tolerancji materia owej znajduj si r wnie w extranecie fi...

Page 25: ...lizacji wielu produkt w w jednym sterylizatorze parowym Nale y dopilnowa aby maksymalny dozwolony za adunek sterylizatora parowego podany przez producenta sterylizatora nie zosta przekroczony Przechow...

Page 26: ...Dystrybutor Aesculap Chifa Sp z o o ul Tysi clecia 14 64 300 Nowy Tomy l Tel 48 61 44 20 100 Faks 48 61 44 23 936 E mail info acp bbraun com TA Nr 012978 11 12 V6 nd Nr 45456...

Page 27: ...valid ci procesu istenia Za to je zodpovedn prev dzkovate osoba vykon vaj ca istenie Pre valid ciu sa pou va doporu en ch mia Ozn menie Ak nenasleduje na z ver steriliz cia mus by pou it virucidn dezi...

Page 28: ...skladujte v obale tesnom proti z rodkom v suchom tmavom a rovnomerne temperovanom pries tore chr nen pred prachom Technick servis Pre servis a opravu sa obr te na svoje n rodn B Braun Aesculap zast pe...

Page 29: ...maddesi kullan lmal d r Not Haz rl k ve malzeme uyumlulu una y nelik g ncel bilgiler i in bak n zAesculap Extranet www extranet bbraun com Do rulanan buharla sterilizasyon prosed r Aesculap Steril Ko...

Page 30: ...vakum s recinde sterilizasyon bekleme s resi 5 dakika Bir buhar sterilizat r nde ayn anda birden fazla r n sterilize edilecekse Buhar sterilizat r n n retici bilgile rine g re azami kapasitesinin a lm...

Reviews: