Software
2
2 - 2
Infusomat® fmS 2.3 gb
Approved Software Versions
IFMC02001
-
Basic software
(Must not be used any more. Please contact the Technical
Service of B. Braun).
IFMC02002
-
Error elimination
at Bolus special function
at staff call on the fm system
-
Optimized air sensor evaluation
IFMC02003
-
EMC optimized
-
New error code FF16, defective membrane keyboard
IFMe02002
-
Preselected volume and time counted down to 0
-
Interface changed to DIANET
Star
(not compatible with Dianet)
-
New special function dose calculation
-
New special function Piggyback
-
New special function clock
-
Storage of alarms in case of malfunctions, which can be re-
called in the Service Program, function 230
IFME03002
Only controller board with loudspeaker
.
Like IFMe02002 and in addition:
-
Alarm volume can be set
-
History function
IFMe02003
-
Dianet Star corrected
IFME03003
-
Additionally History corrected
-
Log of volume delivered
-
New event: normal mode / piggy mode
IFMe02004 / IFME03004
-
Optimized dose calculation
-
New language “Hungarian” in language group E
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...