47
1.
Open the phase control software and click the
icon to open the windows device
manager. Three connected AC Sources should be displayed under ‘Ports (COM & LPT)’ in
the device manager as shown:
If they are missing or shown with an error power off and disconnect the AC Sources from
the computer. Power back on and connect AC Sources to the computer one at a time.
2.
Click ‘Refresh’ and the software will automatically search for connected AC sources and
assign one master and two slaves. Alternatively, you can use the dropdown to manually
assign master and slave AC sources.
3.
Click ‘Connect’ to complete the setup process. A successful connection is indicated by
the green light as shown:
Setting Parameters
Set Voltage (Vrms), Frequency (Hz), and Range (either 150 Vrms or 300 Vrms) for the whole 3-
Phase system.