21
Parameters
Range
Description
List No.
0 to 9
Show configuration of a LIST
Step No.
0 to 99
Show current edit step of list
Volts (rms) Start 0 to 300
Set starting AC voltage
Volts (rms) End
0 to 300
Set ending AC voltage
Volts (DC) Start
-424.0 to 424.0 Set starting DC voltage
Volts (DC) End
-424.0 to 424.0 Set ending DC voltage
Freq. (Hz) Start
43 to 1200.0
Set starting frequency
Freq. (Hz) End
43 to 1200.0
Set ending frequency
Time (ms)
0 to 999999
Set the time interval for running
the specified configuration
CYCLE
0 to 999999
Set the period for running the
specified configuration
Steps
0 to 200
Set the number of steps the
specified configuration will be
divided into
Table 5 - List Config Parameters, Second Page
In the following example 3 new steps will be added to List No. 0.
Figure 17 - List Mode Waveform
Parameters
Step 1
Step 2
Step 3
List No.
0
0
0
Step No.
0
1
2
Volts (rms) Start
80
60
40
Volts (rms) End
0
0
0
Volts (DC) Start
0
0
0
Volts (DC) End
0
0
0
Freq. (Hz) Start
50
50
50
Freq. (Hz) End
50
50
50
Time (ms)
100
60
20
Steps
5
3
2
Table 6 – Step parameters
Summary of Contents for 9830B Series
Page 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Page 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Page 28: ...15 6 Menu Tree ...
Page 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Page 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Page 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Page 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...