SUBJECT TO CHANGE WITHOUT PRIOR NOTICE
5
Before a functional tests, the procedures indicated in this paragraph must be performed for completing the physical installation phases
including the connections.
The device is delivered in
SLEEP
mode. Based on the order, if the installation version and the unit of measurement are already configured,
it is necessary to keep T2 button pressed for three seconds to initialize the device.
5. Commissioning
The first configuration chosen during installation (supply or return pipe) can’t be modified!
If a choice has not been made during the order, the meter must be
configured. The configuration of the meter can be performed via
buttons or via NFC through android app. If the activation takes place
with buttons, the items in the menu will be two:
By pressing for 3 seconds the right T2 button you can confirm the
choice made by passing directly to the initialization of the device.
By pressing the T1 button for 3 seconds the thermal energy meter
will return to installation version mode showing ‘set Mode’ on
display. After confirmation in step 2 the thermal energy meter will
perform the initialization for heating and cooling accounting. The
unit of measurement can be changed later via NFC.
1)
Installation version: select, through the left/right button (T1/T2),
the type of installation (supply or return).
Holding down the right button for 3 seconds can temporarily
confirm the choice made by switching to level 2 (unit of
measurement). If the T1 button is pressed for 3 seconds the
thermal energy meter will return to stock mode showing ‘SLEEP’ on
display.
2)
Units of measurement: select through the left/right button (T1/
T2), one of the following items, ordered as described below:
›
1 (Joule)
›
2 (MJ)
›
3 (GJ)
›
4 (KWh)
›
5 (MWh)
User configuration data can be configured via NFC device and the
BMetering NFC Config app. Below the list of data available:
›
All averages values (temperatures, flow rates, etc) are saved every
hour.
›
The biweekly and monthly historical data, present in level 6, are
saved at each occurrence (default: day 1, end of month). if the
configured days are the same (example: day 15 or leaving the
default values) the data are saved as ‘monthly’ and therefore will
be stored in memory up to 24 previous historical months.
If two different dates are selected (example: day 15 for bimonthly
historical data and day 1 for monthly historical data) a maximum of
48 values are stored in the device (including 24 previous historical
monthly values and 24 previous historical bimonthly values). The
range of days selectable are between 1 and 28.
›
The bi-annual historical data (memory day 1 and 2), present in
level 4, are saved and displayed with the format DD/MM/D1 and
DD/MM/D2 (where Dx indicates the year of saving). Default: 01/09
for memory day 1 and 30/06 for memory day 2. The range of days/
month selectable are between 1..28 (days) and 1..12 (months).