Measuring Blood Oxygen Level with B.O.L.T
1.
Clean the rubber pad of the SpO2 probe touching the finger gently with a soft cloth
dampened with 70% isopropyl alcohol and let air dry or use a soft dry cloth to clean
before measurement.
2.
Clean your hands with hand wash before the measurement.
3.
Select “Check Vitals” option to perform the individual test and continue.
4.
Select the Pulse Oximeter option from the application screen.
Inserting Finger into SpO2 Probe
5.
Clip the SpO2 probe in the index finger of the person to be measured in right/left
hand as displayed above.
6.
Click start/play after inserting the finger inside the SpO2 probe.
7.
Pause to display the SpO2 values.
8.
On successful measurement, the SpO2 value and Pulse rate will be displayed and a
copy of reading will be mailed to your Near & Dear mail IDs.
In Master health chekup, after the successful SpO2 measurement the
application screen will lead to body temperature measurement using Infrared
Thermometer.
Ensure to clean your hands before taking a SpO2 measurement. For
some readings, fluctuations may occur below 85% of SpO2 for certain
ethnicity group.
If the SpO2 values are deviated below 70% or above 100%, ensure
that the user is stable and positions his finger properly as shown below.
This is a continuous measurement procedure. Hence readings SpO2 will
continue to be displayed until the user stops the measurement/test.