ULT FREEZER – USER MANUEL
35
Updated 19. April 2021
6. Set Temperature
To set the setpoint for the temperature in the cabinet.
1.
Use arrow up or down to change the value.
2.
To save the changed setpoint, click on the OK button.
In this example the setpoint is -80,0°C and the Hysteresis is 2,0°C.
That means the controller will start the Compressors at -79,0°C and stop at -81,0°C.