BMF FREEZER – USER MANUAL
32
6. Set Temperature
To set the setpoint for the temperature in the cabinet.
1.
Use arrow up or down to change the value.
2.
To save the changed data click on the OK button.
In this example the setpoint are
-20,0°C and the Hysteresis are 1,0°C.
That means the controller will start the Compressor at -19,5°C
and stop the Compressor at -20,5°C.