1

 - Turn on the scale. Press  

  key to choose the 

weighing

 or 

counting

 functions.

8

 

COUNTING MODE

1

 - After entering into counting mode, press the  

  key to select the sample size.

NOTE

10, 20, 50, 100, 200, 500, 1000 pcs. 

2

 

- Put the sample number on the scale and press  

  key.

When the symbol” c” disappears the sampling procedure is finished and you can start to use the counting 

function.

If the unit weight is too small (less than 0.2 or 0.04d of a scale division), the display will show  “ 

–or – 

pcs

”.

9

 

ACCUMULATION FUNCTION 

In weighing or counting mode, make it to zero, place the goods. 

1

 - After the display is stable, press the  

  key to enter into accumulation mode. 

2

 

After accumulation, total accumulated weight and total accumulation times are shown in display 

window in turn. 

3

 

In accumulation mode, press the

 

 

key to return to operation mode.

Clear accumulation  

Press the  

  key, meanwhile also press the  

  key to clear the memory of accumulation.

Accumulation error ! 

When the total accumulation time is more than 99 times or the total accumulated weight is more than the 
display range, the scale will give you error message, at the same time, the current operation is useless 

!

Summary of Contents for Serenite 9260(A)

Page 1: ...9260 A 9360 b3c pesage pesage w w w b 3 c p e s a g e f r 9260 A 9360 Ver 2014 EN September BC 9260 A 9360 OPERATING MANUAL GB Compacts scales...

Page 2: ...ded for internal operations The scale complies with the type described in the EC type Approval Certificate as well as the appli cable requirement of EC Directives 89 336 EEC and 73 23 EEC SAFETY INSTR...

Page 3: ...f there is permanent electrostatic interference we recommend to connect the platform with the ground The scale is a measuring instrument Air currents vibrations rapid changes in temperature and direct...

Page 4: ...w When the transportation will be made please install the protection screw on the machine to avoid any damage sensitive components 2 LEVEL INDICATOR The scale is equipped with a level indicator on the...

Page 5: ...an press this key the weight will become 0 Tare the weight of a container Indicate the net weight Input sample size To modify the backlight In the calibration mode to change the value 0 9 Accumulate w...

Page 6: ...ht will be effective When the weight is zero for about 8 seconds the backlight will go off automatically bl 2 The backlight will always be on bl 3 The backlight will be turned off 6 TARE FUNCTION Tare...

Page 7: ...ill show or pcs 9 ACCUMULATION FUNCTION In weighing or counting mode make it to zero place the goods 1 After the display is stable press the key to enter into accumulation mode 2 After accumulation to...

Page 8: ...ble press the key Auto calibration is finished WITH AN KNOWN WEIGHT 1 Turn on scale 2 During the scale in self testing press the and keys about 3 seconds 3 When the display shows CC to change the figu...

Page 9: ...ged If it the led is red the battery is nearly discharged If it the led is yellow the battery is increasing the charge level NOTE The battery should changed if the scale is not used for long time 12 T...

Page 10: ...07 31 Load cell 15kg 0 2g 3 4 19020007 32 Load cell 15kg 0 5g 3 5 19020007 41 Load cell 30kg 0 5g 3 6 19020007 42 Load cell 30kg 1g 9 0210002 CPU Main board 10 0210001 LCD Display board 11 0210003 Key...

Page 11: ...ENGLISH 11 3 1 3 2 3 3 3 4 3 5 3 6 1 2 9 10 11 12 3...

Page 12: ...9260 A 9360 ES Balanzas electr nicas 9260 A 9360 IT Bilance electroniche 9260 A 9360 ND Electronische weegschaal Reeks 9260 A 9360 POR Balan as industriais 9260 A 9360 The indicated balances respect t...

Reviews: