The instrument has extremely high stability and reliability, thus is not easy to have error in general situation. Once an error occurs,
please make clear the error first and observe whether the instrument still has error after power-on. Do not hurry to repair the scale
body or instrument. Repair the instrument according to error code of the instrument as possible.
Nº
Symbol
Analysis
Remark
1
[
-EEE
]
[
EEE
]
Unable to reset after startup
1. Determine it is no-load state in startup;
2. Make calibration again.
2
[
]
The weighed object is over full range for 9 days
Reduce weight on weighing platform
3
[
]
The weighed object is below 0 for 5
DAYS
Press
to reset
4
[
]
Out of clearing range
Check whether the weighing platform has weight.
[
]
Remove weight.
5
Invalid operation
6
[
Err 03
]
EEPROM checksum and error
Press
reprint factory value. Start up again. If the
information occurs again, return to factory for repair.
Please calibrate the scale again if the situation does not
occur;
Attention: this place is provided with all parameters of
instruments of the factory.
7
[
Err 0
]
The calibration input weight is too small
Input
≥
10 % weight of full range
8
[
Err 06
]
The weight in calibration is too light
Load
≥
10% weight of full range
9
[
Err 07
]
The scale is dynamic in scale
Inspect the scale body
10
[
Err 08
]
Setting error of date and time
Set according to specification of date and time
11
[
Err 09
]
Error of AD initialization
If the error occurs after restarting, return it to the fac-
tory for repair
12
[
LOAd
]
In loading scale, it indicates to loading weight;
Load weight according to requirements
13
[
Et P
]
It has enters menu setting
Pres [print] to continue setting.
14
[
End
]
End of point and loading point calibration
15
[
Add--
]
Including current display weight in accumulated
value
16
[
-O ER-
]
Accumulated weight overflows
Clear accumulated weight in time.
17
[
Ld---
]
Loading default value
18
[
]
Printing