4. CALIBRATION
PLEASE READ THE ENTIRE CALIBRATION PROCEDURE BEFORE CALIBRATING THE LS8000
.
If no time delay is required, turn the time adjustment pot completely CCW and set S1 down, S2 up. The
potentiometer has 20 turns and freewheels at the end of its travel without damage. You may hear “clicks” at the end
of travel.
A. Setting the Time Delay
1) Setting the timing mode, either on delay or off delay, is done by positioning S3 and S4 as shown in the
chart below.
NOTE:
S3 and S4 also select fail safe high level (FSH) or fail safe low level (FSL).
2) Select the timing range using S1 and S2 as shown on the chart below.
3) Adjust the time adjustment potentiometer clockwise to increase time or CWW to decrease until the exact
time delay is reached. The test switch may be depressed to start the timing when checking timing
sequence.
NOTE:
For long timing ranges you can set the timer for a shorter range, then
change the range switches (S1 and S2) to increase the time. For example, the desired
time delay is 15 minutes. This is equal to 900 seconds. By setting S1 down and S2 up
and adjusting the timer to approximately 28 seconds, the ratio on this range is 1. To
increase this range by 32, move S1 up and S2 up. 32x28 seconds equals approximately
15 minutes.
B. Fail Safe Selection
The fail safe feature provides a “false alarm” in case of power outage or major component failures. When
properly selected, the fail safe feature can protect equipment or alert you of a unit failure.
1) Fail Safe High (FSH) means that the relay is energized when NO PRODUCT is touching the probe. The
green led indicates relay status and is on when the relay is energized. When product comes into contact with
the probe, the relay is de-energized.
2) Fail Safe Low (FSL) means that the relay is energized when PRODUCT IS touching the probe.
NOTE: Due to the wide variety of applications and possible control or alarm functions, the proper fail safe
selection will depend on your circumstances. Please call your distributor or the factory if you need assistance.
TIMING RANGE
FAILSAFE TIMER MODE
S1 S2
TIME
RATIO
S3 S4
DN
UP
1/8 sec - 35 sec
1
UP UP ON FSL
UP
DN
1/2 sec - 140 sec
4
DN UP
DELAY
FSH
UP
UP
4 sec - 18 min
32
DN DN OFF FSL
DN
DN
32 sec - 2 hrs
260
UP DN
DELAY
FSH
C. Sensitivity Calibration
The LS8000 should be calibrated when no material is touching the probe and when the unit is installed where it
will be used.
1) Observe the red LED. If it is on, go to step 3.
2) If the red LED is off, turn the sensitivity adjustment potentiometer CW until it comes on.
3) Turn the sensitivity adjustment pot CCW until the red LED just goes out and continue to turn the pot CCW
at least ¾ of a turn. This sets the unit to its maximum stable sensitivity. If in your application this is too
sensitive, you can turn the sensitivity adjustment pot further CCW. Always bring the product into contact
with the probe after calibration to assure proper performance.
Note the green LED indicates relay status. The relay is energized when the green LED is on.
SPECIAL CALIBRATION NOTES:
For some products, the LS8000 can be calibrated with product touching the probe. To do this, turn the
sensitivity adjust pot CW until the red LED just comes on and continue CW for ¼ to ½ of a turn.
ALWAYS CHECK TO SEE IF THE UNIT RESETS IN THE ABSENCE OF PRODUCT.
10