Begin treatment 

a. If a mouthpiece is used, place the mouthpiece between your teeth. With your mouth closed, inhale deeply and slowly through your mouth as 

the aerosol begins to flow. Hold your breath for several seconds, then exhale slowly through the mouthpiece. 

b. If mask is used, wear the mask and position it over the mouth and nose. As aerosol begins to flow, inhale deeply and slowly through your 

month. Hold your breath for several seconds then exhale slowly. 

14.  After the treatment is complete, switch off the nebulizer by pressing the On/Off switch                 

and then unplug the unit. 

15.   Clean the nebulizer according to the 

“CLEANING” section in the manual. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  Switch off the nebulizer by pressing the On/Off switch 

“A1”. 

  Unplug the nebulizer from the mains. 

  Disconnect the Air Tubing and set it aside. 

  Disassemble mouthpiece parts or mask from the nebulizer set. 

  Open Nebulizer Set by turning the nebulizer cap counter-clockwise. 

  Discard the remaining medication in the nebulizer cup. 

  Wash all items, except tubing and nebulizer, in clean water and non lemon type 

dishwashing detergent solution. Rinse thoroughly with warm water to remove the detergent residue. Air dry or hand dry in a clean environment. 

  When dry, reassemble nebulizer and store the unit. 

  Never use boiling water to clean any parts. 

  Never use hi-temperature sterilizing for disinfection (e.g. baby bottle sterilizer). 

  When needed, only clean the outside of the Air Tubing with dry cloth. Never try to clean the inner wall of the Air Tubing. 

 

 

If the nebulizer is dirty, follow the step below:   

 

Ensure the switch is in the “Off” position and unplug the power cord from the mains. 

  Wipe outside of the nebulizer with a clean, damp cloth. 

  Never use detergent or corrosive substances to clean the nebulizer. 

  It is recommended to clean the nebulizer every few days to keep it clean and dust free. 

 

 

 

  Store the nebulizer and accessories in a clean location. 

  Keep the nebulizer out the reach of unsupervised infants and children. The nebulizer contains small pieces that can be swallowed.   

  Do not store the Air Tube while there is moisture or medication remaining inside it. 

  Clean and disinfect the nebulizer kit, mouthpiece or masks before using them for the first time after purchase. Also clean and disinfect the 

nebulizer kit if the device has not been used for a long period of time. 

 
 
 
 
 

 

 

Always disinfect the nebulizer set, mouthpiece and other nebulizer parts before using the nebulizer for the first time, after the last treatment of the 
day, or daily during normal use. 

  Disconnect the Air Tubing and set aside. 

  Clean the casing of the nebulizer and air connectors by using a soft cloth moistened with alcohol. Do not use abrasive cleaners. Dry the casing 

immediately using a soft clean cloth. 

  Disassemble mouthpiece parts or mask from the nebulizer accessories set 

  Using a clean container or bowl, soak all accessories item except Air Tubing in 75% alcohol or 3% hydrogen peroxide for 5 minutes or use a 

medical bactericidal disinfectant available through your provider. 

 

 

  With clean hands, remove the accessories item from the disinfectant solution, rinse under clean water. 

  Air dry or hand dry in a clean environment. 

  When dry, reassemble the nebulizer and store for the next use. 

 

 

The nebulizer has an Air Inlet filter to remove airborne contaminants. On average, the Air filter requires regular replacement after every 60days. The 
nebulizer comes with a package of extra filters for this purpose. Check the Air Inlet filter at least once every two weeks. Following the below 
mentioned steps to check and replace the Air Filter when needed.   

1. Hold the Air Inlet filter cover 

”A3” with your fingers and turn anti -clockwise to unlock it and then pull out. 

2. Now you can see the Air Filter in the filter compartment. Remove the Air Filter from the filter compartment (if needed, you may use a toothpick to 

remove the old filter).   

 

 

3. If change is required, insert a new Air Filter into the filter compartment and discard the dirty Air Filter, otherwise reinsert the original Air Filter into 

the filter compartment. 

4. Insert the filter cover in position and turn clockwise to lock it in position. 

 

 

1. If the cord is damaged, it must be replaced by an Authorized Service Centre.   

 

 

 

2. If the nebulizer fails to operate, unplug the unit for 15minutes, plug in and try again. If the problem persists, check the 

“TROUBLE SHOOTING” 

section. If problem still persists, then contact Authorized Service Centre. 

 

 

Nebulizer not 
work at all   

Check that the power cord is connected to the nebulizer and 
plugged in the mains. Check the fuse in the plug if there is any. 

Nebulizer is 
noisy 

Check that the filter is locked correctly 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLEANING THE NEBULIZER SET

 

NOTE: The compressor uses an automatic reset thermal limiter that shuts off the 
nebulizer if the motor is overheated due to misuse, abuse or abnormal operation.   

The nebulizer will restart when the motor cools to an acceptable level. 

 

NOTE: the Air Filter should be changed if significant dust is noticed around the middle of 
the Air Filter.   

 

 

 

 

NOTE: Alcohol is highly flammable; do not use alcohol in close vincinity of open fire; do 
not smoke. 

 

 

 

 

CAUTION: To prevent electric shock, DO NOT DISASSEMBLE and never immerse   
the nebulizer in water or other liquid or place it under running water. Always disconnect 
the unit from mains before you start to clean it. It is recommended to clean all parts 
(except the air tubing and nebulizer) after every use. 

 

 

REPLACING THE FILTER

 

NEBULIZER CLEANING

 

TROUBLESHOOTING   

SERVICE AND CUSTOMER CARE 

DISINFECTION

 

STORAGE 

 

NOTE: Special care is needed to prevent spilling the medication when attaching the Air 
Tube (i.e. keep the nebulizer kit upright). 

 