Generally, accepted “normal” temperature reading is 37.0°C /98.6°F, but 
temperature readings differ to this figure may still be considered “normal”.   
Variations in temperature can be attributed to activities such as exercise, 
smoking, eating and drinking. Our temperature is lower in the morning than in 
the afternoon. Other variations may be due to the location of the temperature 
reading.  
Method of measuring also varies the measured temperature as well and normal 
temperature ranges are show as follows: 

Oral 35.5-37.5˚C (95.9-99.5˚F) 

Axillary (underarm) 

34.7-37.3˚C (94.5-99.1˚F) 

Rectal 36.6-38.0˚C (97.9-100.4˚F) 

Ear 35.8-38.0˚C (96.4-100.4˚F) 

 

When use the thermometer first time, open the battery cover as per the 
“REPLACING THE BATTERY” section in this manual.    Put 2 AAA (LR03) 
batteries in the battery compartment with correct polarity. Replace the cover and 
lock.    It will switch ON automatically in the Stand-By Mode.    The default time 
setting is 12:00. 

The thermometer can also be changed to Temperature Measure Mode.   
Switching to this mode, the ear canal temperature and the forehead 
temperature can be measured. 

Details of operation in each mode are described in the following sections. 

 

 

  In the Stand-By Mode, the time display and the ambient temperature display 

are shown on the LCD display screen alternately.    The time display shows 
the clock in hh:mm (24-hour) format.    The ambient temperature display 
shows the measured room temperature in temperature unit (°C or °F) that is 
last used in the Temperature Measure Mode. 

  The time display and the ambient temperature display interchange every 20 

seconds.    If the thermometer is in the Temperature Measure Mode, it will 
switch back to Stand-By Mode after left idle for 30 seconds. 

 

 

  In the Stand-By Mode, wait the display screen to show the time display. 

  Press and hold Button A for 2 seconds, the hour digits flash.    Press Button B 

to adjust the hour by advancing one hour. 

  Press Button A again will set the hour and swap the minute digits to flash.   

Press Button B to adjust the minute by advancing one minute.    Keep 
pressing the Button B can increase the advancing pace at high speed. 

  Press Button A will interchange the adjustable unit (hour & minute).    Press 

and hold Button A will complete the time setting process.  Stand-By Mode 
resumes. 

 

  This thermometer is designed to receive the infra-red rays emitted by the skin 

surface and in the ear drum. It provides a fast and accurate means for you to 
check your baby’s temperature easily. 

  If the thermometer is in the Stand-By Mode, press Button B to change to 

Temperature Measure Mode.    The thermometer will briefly show all icons on 
the display as per the image of “Fig. A”, and then thermometer will show 

0.0˚C or 0.0˚F, this indicates the thermometer is ready for measure 
temperatures.    The measuring method (ear or forehead) follows the usage in 
the last temperature measurement. 

  Press and hold Button A will interchange the measuring method between ear 

and forehead. 

  Press and hold Button B will interchange the temperature unit between °C 

and °F 

 

 

Fig. B 

                       

 

  Take off the protective cover, change thermometer to Temperature Measure 

Mode.    Select the measuring method to ear measurement and the preferred 
temperature unit (°C or °F). 

  Insert the probe into the ear canal gently as per “Fig. B”. 

  Press Button B.    The thermometer responses with a beep and it will 

automatically read the ear temperature in around 2-3 seconds. 

  Once the measurement completed, the thermometer will beep again. 

  Measured temperature will then show on the display. 

Note:

   

1. Please ensure the probe is inserted into ear canal to get an accurate 

temperature measurement. 

2. For optimal precision, ensure the ear is not obstructed by wax or a plug. 
3. As usual for any measuring equipment, slight variations in temperature may 

results for the measurement one after another.    It is therefore recommended 
to take several readings and take the highest in account (Withdraw the 
thermometer from the pinna and press Button B briefly between each 
reading).    Measure again and new measured temperature will then show on 
the display within another 2~3 seconds.   

 

Fig. C 

                       

 

  Take off the protective cover, change thermometer to Temperature Measure 

Mode.    Select the measuring method to forehead measurement and the 
preferred temperature unit (°C or °F)   

  Place the probe to the middle of the forehead (just above the surface of the 

skin) with the probe perpendicular to the skin as per “Fig. C” 

  Press Button B.    The thermometer responses with a beep and it will take the 

forehead temperature in around 2-3 seconds. 

  Once the measurement completed, the thermometer will beep again. 

  Measured temperature will then show on the display.   

 

  If measured temperature within normal range, thermometer will beep once 

and then the peak temperature measured with a smiling face icon “

” will 

show on the display to indicate normal temperature is measured.   

  Otherwise, if the measured temperature is above 38.0°C (100.4°F), the 

thermometer will show an sad face icon “

” with the peak measured 

temperature on the display, indicating a fever signal.    In this case, you 
should consult your physician. 

  Read and record the temperature after each measurement.    Last 12 

readings will be automatically stored in the memory for reference; 

  The thermometer will switch off automatically in 60 seconds if no button is 

pressed again after it is activated by pressing the “Power On” button.   

 

 

  This thermometer can store last 12 temperature measurements.   

  To recall the memories, press Button A in Temperature Measurement Mode. 

The memory icon “

” and a number will show on the display.    The number 

show on the display represents the temperature recorded in the memories. 
“0” is the last memory and “1” is the second last and so on.    Press 
successively to scroll the memories.     

 

1. The Lens is the most delicate part of the unit, and special care is needed   
2. Clean the thermometer using soft dry cloth or cotton swab.   

 

The battery in your infra-red thermometer needs to be replaced when “

” 

appears on the display window.   
To replace the battery     
1. Remove the battery cover by turning the battery cover lock with a coin in 

anti-clockwise direction. 

2. Take out the old batteries from the thermometer and replace with the new one, 

Ensure to fit the batteries with correct polarity. (Note: polarity mark of the 
batteries are provided on the battery compartment for easy reference)   

3. Replace the battery cover and turning the battery cover lock back in place. 

NOTE: Always use environmental friendly way to dispose 
batteries.

 

  If measured temperature does not seem to be in normal range or unstable, 

you should consult your doctor or physician to check your health condition. 

 

 

Dimension (mm) 

131.5(L) x 35.5(W) x 30(D) 

Weight 

70 grams (including batteries) 

Measuring time 

As short as 2-3 second.    . 

Measurement 
range and 
Precision 

 
+/-0.2°C (0.4°F) for 36 - 39 °C (96.8 - 102.2 °F) 
+/-0.3°C (0.5°F) for T <36°C (96.8°F), or >39°C (102.2°F) 
 

Measurement 
units 

Measure in Centigrade (°C) or Fahrenheit (°F)     

Display 
Resolution 

0.1 for both °C and °F     

Battery 

2 x LR03 /AAA size 1.5V batteries or equivalent 

Operating 
Condition 

Temperature: 16 to 40 °C (60.8 to 104 °F) 
Relative Humidity: < 85%   

Storage Condition

Temperature:

10 to 55 °C (–14 to 131 °F) 

Relative Humidity: < 85% 

Other feature 

With a lanyard hole for easy carrying and storage 

This technical specification may subject to modification to improve the 
thermometer, without prior notice. 

 

 

FOREHEAD MEASUREMENT

HOW TO SET TIME

 

Stand-By Mode

 

SPECIAL NOTE 

 

BEFORE FIRST USE

 

WHAT IS A “NORMAL” TEMPERATURE?

AFTER YOU GET THE READING

EAR MEASUREMENT

 

PRODUCT FEATURE & SPECIFICATION

 

REPLACING THE BATTERY   

TEMPERATURE MEASURE MODE

 

RECALL THE MEASUREMENT STORED 

 

GENERAL CARE 

 

DO NOT hold the thermometer too 
long, as it may affect and render 
the measured temperature to lower 
than usual.   

Reviews: