13
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity (For home healthcare environment)
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrostatic transient/burst
IEC 61000-4-4
± 2 kV ,
100kHz, for
AC power port
± 2 kV ,
100kHz, for
AC power port
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5
±0.5kV, ±1kV
(differentia mode)
±0.5kV, ±1kV
(differentia mode)
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°
, 270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°
, 270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
30 A/m; 50Hz or 60Hz
30 A/m; 50Hz or 60Hz
Power frequency
magnetic fields should be
at levels charactertic of a
typical location in a typical
commercial or hospital
environment.
Summary of Contents for NURSE PRO
Page 30: ...30 BG NURSE PRO LD 202 1 2 3 BF 20 o C 55 o C 4 o F 131 o F 15 93 EC REP...
Page 32: ...32 5 6 5 1 5 2 Figure 5 3 Figure 5 1 Figure 5 2 5 3 5 4 1 2 3 4 6 1 6 2 6 3 6 4...
Page 34: ...34 4 1 5 LCD...
Page 35: ...35 a 1 2 3 5 4 6 7 6 1 2 3 4 7 5...
Page 36: ...36 30 24mm 27mm...
Page 37: ...37 1 2 3 4 5 1 2 3 5 4 5 1 2 3 5 min Err...
Page 38: ...38 1 2 3 4 5 6...
Page 41: ...41 IEC 60601 1 2 HF 1 CISPR 11 1 CISPR 11 1 IEC 61000 3 2 Class A IEC 61000 3 3...
Page 184: ...184 RU NURSE PRO LD 202 1 2 3 BF 20 o C 55 o C 4 o F 131 o F 15 93 EC REP...
Page 186: ...186 5 U 6 Figure 5 2 5 3 5 4 5 1 5 2 Figure 5 3 1 2 3 4 6 1 6 2 6 3 6 4...
Page 187: ...187 4 1 5...
Page 188: ...188 a 1 2 3 5 4 6 7 5 6 1 2 3 4 7...
Page 189: ...189 30 24mm 27mm...
Page 190: ...190 1 2 3 4 5 1 2 3 5 4 5 1 2 3 5 min Err...
Page 191: ...191 1 2 3 4 5 6...
Page 194: ...194 IEC 60601 1 2 HF 1 CISPR 11 1 B CISPR 11 1 B IEC 61000 3 2 A IEC 61000 3 3...
Page 226: ...226 UA NURSE PRO LD 202 1 2 3 BF 20 o C 55 o C 4 o F 131 o F 15 93 EC REP...
Page 228: ...228 5 6 5 1 5 2 Figure 5 3 Figure 5 2 5 3 5 4 1 2 3 4 6 1 6 2 6 3 6 4...
Page 229: ...229 4 1 5...
Page 230: ...230 a 1 2 3 5 4 6 7 5 6 1 2 3 4 7...
Page 231: ...231 30 24mm 27mm...
Page 232: ...232 1 2 3 4 5 1 2 3 5 4 5 1 2 3 5 min Err...
Page 233: ...233 1 2 3 4 5 6...
Page 236: ...236 IEC 60601 1 2 HF 1 CISPR 11 1 B CISPR 11 1 B IEC 61000 3 2 A IEC 61000 3 3...
Page 239: ...239 80 800 80 2 7 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3 8 7 3 100 12 23 d P 1 80 800 2 4...
Page 240: ......
Page 242: ......
Page 243: ......
Page 244: ...Cat No 1483...