4.3. Combustion Testing Process
WARNING: The Fyrite
InTech
calculates combustion parameters
based on North American or Siegert combustion equations. NA or
Siegert configuration is selected in the SETUP MENU. Be sure that
your Fyrite
InTech
is properly configured for your region and
desired combustion calculations.
NOTE: The recommended time required to achieve a stable
measurement is a minimum of 3 minutes.
Step
Example Combustion Testing Procedure
1
Confirm that testing prerequisites have been completed.
2
Based on the sampling point examples and your particular combustion
application, locate and prepare an appropriate sampling point.
3
Insert the probe into the combustion location.
4
Press the RUN/HOLD button to begin sampling gas. You should see the
word RUN in the upper left corner of the display and hear the sample pump
turn on. If you see the word HOLD, press the RUN/HOLD button again.
5
Monitor the display for combustion data.
6
If desired, turn on your optional IrDA printer, then press the F1 button on
the
Fyrite
InTech
to print the current combustion data.
7
Press the F3 button as desired to save combustion data for later
retrieval, review, and/or printing.
8
Press the RUN/HOLD button to stop the test. You should see the word
HOLD in the upper left corner of the display and hear the sample pump
turn OFF. If you see the word RUN, press the RUN/HOLD button again.
9
Remove the probe from the sampling point.
CAUTION: The probe may be very hot. Allow it to cool, then
wipe it clean with a dry cloth.
10
Move the instrument to a clean air environment and press the POWER
button to turn off the instrument. The shutdown procedure includes a
purge component that clears the sensors of combustion gases.
11
Turn on the instrument to optionally print and/or evaluate saved test