Operation
Fyrite
InTech
Manual
48
0024-9486 Rev 4
Fyrite
InTech
can be setup to include a custom logo on printouts. Logos are
loaded into the instrument using the Fyrite
User Software (FUS). Logo size is
limited to 192 x 384 pixels (height x width) and must be in one of the following
formats: .BMP, .JPG, .PNG, or .TIFF. For best results, the logo should be saved
in black and white.
4.6. Taking Ambient CO Measurements
This procedure takes approximately 15 minutes to complete and provides a
minute-by-minute snapshot of CO readings, as well as a “Max CO” value that
represents the highest CO reading measured during the entire 15-minute test.
Results can be saved to memory and/or printed. Use the following procedure to
perform an ambient CO measurement.
Step
Example Procedure for Taking Ambient CO Measurements
1
Turn on the InTech
in a fresh air environment and wait for
initialization to complete.
2
Verify successful initialization (no errors).
3
Check battery status (see page 15). If battery life is questionable,
replace the batteries, as the Ambient CO test takes approximately
15 minutes to complete.
4
Move instrument to target location to be tested.
5
Press F2 to display the Main Menu.
6
Use the down arrow to highlight Ambient CO Test and press the
ENTER button.
7
Follow the on-screen instructions to initiate the test.
8
Refer to page 20 for details on navigating the ambient CO test
screens, viewing results, saving results to memory, and printing
results.