Instruction 0024-9435
28
Fyrite Tech
5.2 Ambient Temperature Calibration
Material Required:
Calibrated Thermometer
Procedure:
1. Enter the Calibration Mode as described in Section 5.1. Observe that
"AMBIENT" is displayed at the top of the display; if not, repeatedly
press
ENTER
until it appears.
2. Use the and buttons to set the displayed value to match the
reading of a calibrated thermometer at room temperature.
3. Press
ENTER
to store the displayed value and move to the next cali-
bration screen, or hold down
ENTER
for 2 seconds to store the dis-
played value and enter the Run Mode, or press the
I/O
button to exit
the Calibration Mode and turn OFF the analyzer without saving the
changes.
5.3 Stack Thermocouple Channel Calibration
Material Required:
Thermocouple
Simulator
Procedure:
1. Attach the thermocouple simulator to the Stack Ther mo cou ple con-
nector on the analyzer (T-STACK, see Figure 3-2).
2. Enter the Calibration Mode as described in Section 5.1. Then repeat-
edly press the
ENTER
button until “STACK” appears in the top of
the display.