H25-IR PRO Instruction Manual
Wiring
P/N: 3015-5678 Rev 6
33
Table 4-2. Procedure to Change Factory-Set Probe Configuration
Step
Description
1
Remove power cord from back panel.
2
Remove both the front and rear bezels by first disengaging the plastic tabs on the right and left sides of the
bezel, and then pulling the bezel straight out.
3
Remove the top two screws that secure the front panel, and the top two screws that secure the rear panel.
4
Find the seam at the side of the enclosure nearest the top. Using your fingers gently pry up on one side of the
enclosure lid. The lid will disengage and swing upward, leaving the opposite side of the lid engaged. With the lid
freely moving in the opposite side, gently lift up the front and rear edges of the lid to fully disengage it. Set it to
one side and continue with the replacement.
5
Connect the ribbon cable from the front panel PROBE connector to either J2 or J7 according to the probe to be
connected:
•
J7 – Standard probe with LED and push button (and Standard probe with flex tubing)
•
J2 – Smart probe
IMPORTANT:
Note the orientation of Pin 1 on both the ribbon cable and connectors J2 and J7.
The red stripe on the ribbon cable denotes pin 1.
6
Reassemble the case, then connect the power cord.
Figure 4-3. Front Panel Printed Circuit Board Containing Probe Connectors J2 and J7
Ñ
Ñ
Ñ
Summary of Contents for H25-IR PRO
Page 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Page 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Page 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Page 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Page 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...