H25-IR PRO Instruction Manual
Maintenance
P/N: 3015-5678 Rev 6
89
Figure 6-1. Location of Fuse Holder
6.2. Probe Tip Filter Replacement
Table 6-3. Items Required for Probe Tip Filter Replacement
Item
Description
1
Probe Tip Filter (diameter: 0.287 in. [7.2 mm]) (Available in bags of 50 from Bacharach. Refer to Probe Tip Filter
Replacement on page 89)
2
Pin or tweezers
Table 6-4. Probe Tip Filter Replacement Procedure
Step
Description
1
Unscrew metal tip from flexible probe.
2
Remove filter from inside metal tip using pin or tweezers.
3
Insert new filter into probe tip.
4
Inspect “O" ring. Replace “O" ring if defective. (See Probe “O" Ring Replacement on page 90 for information.)
5
Replace metal tip onto flexible probe.