background image

WARNING:

 The MGS-550 MUST NOT

  be  in  an  alarm  or  fault 

condition during calibration. Acknowledge any alarms or faults 

BEFORE attempting to begin the calibration process.

WARNING:

 Except for CO

2

 or O

2

 sensors, calibration gas must be 

in a balance of air, not nitrogen (N

2

).

IMPORTANT:

  Calibration  and  /  or  bump  testing  requires  the 

MGS-550 calibration adapter kit (P/N: 1100-0034).

1.  Ensure the instrument has been powered on for at least 1 

hour prior to beginning the sensor adjustment procedure. 

For semiconductor sensors, you must wait at least 24 hours.

2.  Assemble the calibration kit.
3.  Firmly  push  the  calibration  adapter  over  the  calibration 

port on the sensor assembly.

4.  If  using  a  variable  flow  regulator,  adjust  the  gas  flow  to 

approximately 0.3 L/min.

6b.  Zero Adjustment

WARNING:

 Except for CO

2

 or O

2

 sensors, ambient air may be used 

instead of zero gas if the area is know to be free of the target gas 

or any gases to which the sensor may be cross-sensitive.

5.  Enter the menu and select function F-02 Zero Sensor:

 

¾

Press the [UP]/[DOWN] arrows to select the sensor to be 

calibrated, then press [

].

6.  Apply Synthetic air 

(i.e., 20.9% zero air) or nitrogen, 

and then 

press [

].  

 

¾

T

he  current  value  will  be  blinking  in  the  display  (e.g., 

“2”). The gas value may go negative. This is normal.

7.  Wait for the current gas concentration value to stabilize.
8.  Press [

↵]

 to save the current calibration.

 

¾

The  instrument  will  indicate  success  status  with  a 

“PASS” display. If a “FAIL” message is displayed.

9.  Turn off gas flow and remove the calibration adapter from 

the sensor or disconnect the tubing.  

 

¾

For  oxygen,  ensure  that  the  concentration  is  below 

the enrichment alarm thresholds. 

 

 

Refer  to  user  manual  (P/N  1100-1000)  for 

troubleshooting.

10.  Press [

] to exit the function

11.  Enter the menu and select function F-03 Zero Sensor:

 

¾

Press  the  [UP]/[DOWN]  arrows  to  select  the  sensor 

to  be  calibrated,  then  press  [

].  The  last  span  gas 

concentration will be displayed

12.  Apply calibration gas at the concentration listed on the cal gas 

concentration label 

(located on top of the instrument)

:

 

•  50% Max Range for Butane, CO, CO2, Methane, NH3, NO2, O2, and Propane 
(0-100% LEL) gas detectors

 

•  Up to 100% (do not exceed 100%) Max Range for all other gas detectors

Calibration 

Adapter

7.  Bump Test

1

-Each of the three relays can be configured for low alarm, medium alarm, high alarm or fault – this example of operation assumes a low alarm, high alarm 

and fault configuration

5.  Operation Overview

Button / Mag Switch

Description

Hold:

 3sec - Access User Menu

Hold: 

3sec - Returns to next higher menu without saving changes (ESC)

Press / Tap: 

Brings up sensor ID in succession

  

 

Press / Tap:

 Scrolls through User Menu in sequence

Hold: 

Accelerates scrolling

Hold Up Arrow at Top of Menu: 

Enters Diagnostic Screen

Press / Tap:

 Confirms menu selection

Press / Tap During Alarm:

 Mutes buzzer

Hold:

 Cancels Off-Line Mode

Hold: 

Acknowledge latched alarm / fault

STATE

LEDs

Display

Relay 1

1

Relay 2

1

Relay 3

1

Buzzer

Warm-Up

Flashing Green, 0.5 Hz

Display is blank

Off-line

Off-line

Off-line

OFF

Normal

Constant Green

Toggles between target gas name and real-time 

concentration for each sensor installed

Normal

Normal

Normal

OFF

Low Alarm 

(Sensor 1 and / or 

Sensor 2)

Flashing Red, 0.5 Hz

Display will toggle between ALARM, gas name, 

concentration of each sensor until concentration falls below 

low alarm level

Alarm 

State

Normal

Normal

PULSE, 0.5 Hz

High Alarm 

(Sensor 1 and / or 

Sensor 2)

Flashing Red, 2 Hz

Display will toggle between ALARM, gas name, 

concentration of each sensor until alarm is acknowledged 

and concentration falls below High alarm level

Alarm 

State

Alarm 

State

Normal

PULSE, 2 Hz

Fault Alarm 

(Sensor 1 and / or 

Sensor 2)

Constant Amber

Display will toggle between FAULT 1 (or 2), gas 

name, concentration of each sensor until the fault is 

corrected and acknowledged

Normal

Normal

Fault 

State

CONSTANT

Off-line

Green > Amber > Red

“oFFLn” is displayed

Off-line

Off-line

Off-line

OFF

1.  Connect  adapter  and  gas  cylinder  according  to  the 

instructions in the General Calibration Procedure.

2.  If  desired,  disable  /  silence  external  annunciators 

(e.g., 

shutdown valves, notification of authorities, etc.)

:

3. Apply a sufficiently high concentration of the target gas to 

trigger alarms, but NOT pure refrigerant or hydrocarbons 

(e.g., do not use a butane lighter)

.

4. Once thresholds have been exceeded, relays should activate, 

digital outputs should transmit the gas concentration and:

 

¾

Gas concentration should be displayed, the instrument 

status  should  be  “Low  Alarm”  or  “High  Alarm”  and 

alarms states should be “On.”

5. Turn off gas flow and remove the calibration adapter.
6.  Allow  sensor  to  recover  /  stabilize  before  the  instrument 

returns to normal operation 

(green LED)

.

4.  Installation 

(Continued)

PN:   1100-XXXX

SN:   YYMM####

GAS:   CO2

 

10000 PPM

Part Number

Serial Number

Sensor Type

Max Range

IMPORTANT:

 Both tactile (Press or Hold) and magnetic (Tap or 

Hold) switches are provided depending on the users method of 

interface.  Both  sets  of  switches  have  redundant  functions  and 

can be used to configure, calibrate and maintain the MGS-550. 

The following table describes the switches functionality:

STEP 4 | 

Reconnect Lid 

(Continued)

1.  Carefully  reattach  the  ribbon  cable  to  J1  as  shown  in 

Internal Wiring Description in Section 2, carefully noting 

the keyed position on the connector.

2.  Using a 5/32″ 

(4 mm)

 hex key / allen wrench, tighten the 

lid screws in an “X” tightening pattern.

IMPORTANT:

 If there was a sensor element installed in Step 3, 

the sensor needs to be registered with the gas detector.

1.  Power up the gas detector. The detector will prompt you to 

start the registration. (e.g., “rEg 2”).

2.  Acknowledge by tapping [

]. The detector will indicate 

a successful registration (e.g., “PASS2”) if the sensor was 

correctly installed.

STEP 5 | 

Register the New Sensor

6.  General Calibration Procedure

6c.  Span Adjustment

13.

 

Press the [UP] / [DOWN] arrows to change the values to match 

the concentration of the calibration gas (in the units of the 

sensor – i.e., ppm, Vol %, %LEL, etc.) and then press [

].

14.  Apply span gas and then press [

].

 

¾

The  display  will  show  the  current  gas  concentration 

value blinking in the units of the sensor – i.e., ppm, Vol 

%, %LEL, etc. (e.g., “100”).

15.  Wait for the current gas concentration value to stabilize.
16.  Tap [

] to perform the calibration adjustment.

 

¾

(The  instrument  will  indicate  success  status  with  a 

“PASS” display. Otherwise a “FAIL” message is displayed.)

17.  Turn off gas flow and remove the calibration adapter from 

the sensor or disconnect the tubing.

18.  Ensure  that  the  concentration  is  below  the  alarm 

thresholds. 

 

¾

(For oxygen, ensure that the concentration is below the 

enrichment alarm thresholds.)

19.  Tap [

] to exit the function.

IMPORTANT:

  Refer  to  MGS-550  User  Manual  for  complete 

configuration and operation instructions.

Summary of Contents for MGS 550

Page 1: ...Fixed Gas Detection P N 1100 0618 April 2021 Revision 0 Quick Start Guide Dual Sensor Gas Detector for Refrigerant Toxic Combustible Gas Detection ...

Page 2: ... professional before connecting the MGS 550 to devices not mentioned in this manual Failure to comply may result in injury and or damage to the product CAUTION Except for the maintenance detailed in this manual this product should ONLY be opened and or serviced by authorized personnel Failure to comply may void the warranty 1 Introduction The MGS 550 gas detector is designed for use in refrigerati...

Page 3: ...in their particular industry country This document is only intended as a guide and the manufacturer bears no responsibility for the installation or operation of this unit IMPORTANT Comply with all local and national laws rules and regulations associated with this equipment Operators should be aware of the regulations and standards in their industry locality for the operation of the MGS 550 ...

Page 4: ...Block 1 Pin 10 Relay Wiring Terminal Block T2 Front Panel Component Description 1 Controller Power LED 2 Concentration LEDs PPM LEL Vol both off 3 5 digit 7 segment display Gas real time concentration 4 Magnetic Switches dashed line around icon 5 Tactile Switches solid line around icon 6 Local gas sensor element remote sensor elements available 3 Component Overview ...

Page 5: ...he application possible leaks air movement draft etc when selecting a location Accessibility the degree of accessibility required for maintenance purposes when selecting a location Target Gas the specific gravity of the target gas when selecting the height of the instrument 2 Using a 5 32 4 mm hex key Allen wrench not included remove the lid and disconnect the ribbon cable from the base 3 Set the ...

Page 6: ...n or b an appropriately rated UL listed CSA Class 2 transformer IMPORTANT Refer to MGS 550 User Manual for instructions for connecting to the MGS 402 or MGS 408 controller 1 Strip 0 2 to 0 25 inches 5 to 7 mm of wiring insulation 2 Unscrew the appropriate cable gland and gasket 3 Connect the wires as indicated in the Internal Connections in Section 3 Component Overview Refer to the MGS 550 User Ma...

Page 7: ...gram below If both sensors are to be installed none installed from the factory follow the above instructions for each sensor to be installed WARNING DO NOT leave excess cable inside of the gas detector housing Failure to comply may result in damage to the product CAUTION When installing the sensor ribbon cable care must be taken to ensure the proper orientation of the connector at both ends of the...

Page 8: ...ontinued 1 Carefully reattach the ribbon cable to J1 as shown in Internal Wiring Description in Section 2 carefully noting the keyed position on the connector 2 Using a 5 32 4 mm hex key allen wrench tighten the lid screws in an X tightening pattern ...

Page 9: ... registered with the gas detector 1 Power up the gas detector The detector will prompt you to start the registration e g rEg 2 2 Acknowledge by tapping The detector will indicate a successful registration e g PASS2 if the sensor was correctly installed STEP 5 Register the New Sensor ...

Page 10: ...d or Sensor 2 Flashing Red 0 5 Hz Display will toggle between ALARM gas name concentration of each sensor until concentration falls below low alarm level Alarm State Normal Normal PULSE 0 5 Hz High Alarm Sensor 1 and or Sensor 2 Flashing Red 2 Hz Display will toggle between ALARM gas name concentration of each sensor until alarm is acknowledged and concentration falls below High alarm level Alarm ...

Page 11: ...or CO2 or O2 sensors ambient air may be used instead of zero gas if the area is know to be free of the target gas or any gases to which the sensor may be cross sensitive 5 Enter the menu and select function F 02 Zero Sensor Press the UP DOWN arrows to select the sensor to be calibrated then press 6 Apply Synthetic air i e 20 9 zero air or nitrogen and then press The current value will be blinking ...

Page 12: ...should be displayed the instrument status should be Low Alarm or High Alarm and alarms states should be On 5 Turn off gas flow and remove the calibration adapter 6 Allow sensor to recover stabilize before the instrument returns to normal operation green LED PN 1100 XXXX SN YYMM GAS CO2 10000 PPM Part Number Serial Number Sensor Type Max Range 6c Span Adjustment 13 Press the UP DOWN arrows to chang...

Page 13: ...A Customer Service 1 724 334 5000 CAN Customer Service 1 905 882 8985 EU Customer Service 353 1 284 6388 website mybacharach com email help mybacharach com Copyright 2021 Bacharach Inc All Rights Reserved ...

Reviews: