5.4 Calibrate TA-Zero
Material Required:
• Thermocouple Simulator (K-type)
Range: 0 to 300°C
Accuracy: ±0.3°C
Procedure:
1. With the analyzer turned off, first plug the simulator's K-type connec-
tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.
2. Enter the Calibration Menu Screen per Section 5.3; then choose
TA-ZERO to zero the analyzer’s stack temperature channel.
3. Adjust the simulator to 0° C (32° F).*
4. Wait until the MEASURED reading on the screen stabilizes. Then use
the
st
and
keys to enter an APPLIED value that equals 0 °C
(32 °F).
Typical Calibrate TA-Zero Screen During Calibration Procedure:
CALIBRATE TA-ZERO
MEASURED 3.0°C
APPLIED 0000.0°C
4. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 0 – 5 °C (32 – 41°F). Any
attempt to calibrate outside this range will cause the analyzer to
display an error message.
Calibration
5-4