Programming
Page 24
May 2020
DSY-PM-01378-EN-08
PROGRAMMING
The order of the following programming parameters assumes the meter is set for liquid�
Parameters for gaseous fluids can be found in
OTE:
N
All of the following parameters appear in
Extended Programming
mode�
Parameters with an asterisk (*) appear in
Programming
mode as well�
Saving Programmed Parameters
When navigating through the menus with the
ENTER
button, the programmed parameters
are not permanently saved� Make sure to use the
MENU
button to save that information
before leaving the meter�
IMPORTANT
If the menu times out before the parameters are saved with the
MENU
button, all that
programming information is lost. Make sure to save with the
MENU
button.
Liquid
Select Fluid*
At the
Fluid
prompt, press
ENTER
to view the current fluid type� If the current fluid type is
correct, press
ENTER
to advance to the next parameter� To change the fluid type, press
UP
or
RIGHT
to switch between
Liquid
or
Gas
� Press
ENTER
to advance to the
Meter
parameter�
Select Meter Size*
At the
Meter
prompt, press
ENTER
to display the current meter size� If the current meter
size is correct, press
ENTER
to advance to the next parameter� To change the meter size,
press
UP
or
RIGHT
to scroll to the correct meter size� Press
ENTER
to advance to the
next parameter�
OTE:
N
The meter size selection refers to the bore of the meter and not the connections
size� For a listing of the Blancett turbine bore sizes, see the default K-factor table
in
“Default K-Factor Values” on page 39
�
OTE:
N
In
Programming
mode, the monitor advances to the
KFacUnit
parameter�
See