Programming
Page 37
May 2020
DSY-PM-01378-EN-08
At the
TripSP1
prompt, press
ENTER
� The current tripping condition setting displays� If the
current setting is correct, press
ENTER
to advance to the next parameter� If the current
setting requires a change, press
UP
or
RIGHT
to change to the alternate choice� Press
ENTER
to advance to the
SetPt 2
parameter�
The
SetPt 2
,
HystSP2
and
TripSP2
parameters are set using the same procedures as the
SetPt 1
,
HystSP1
and
TripSP1
parameters� When these parameters have been entered, the
monitor advances to the
Clr G-T
parameter�
Clear Grand Total
At the
Clr G-T
prompt, press
ENTER
� The monitor displays
No
on the screen� To clear the
grand total, press
UP
or
RIGHT
to change from
No
to
Yes
� Press
ENTER
to advance to the
Passwd
parameter�
The totalizer can also be reset using a hardware reset, as shown in
or by pressing
MENU
and
ENTER
simultaneously�
TB1
Mag
Pulse
Iso
OC
Low
High
Input
Total P
ulse
Sig
nal
Freq. In
4-20mA
Iso Total Pluse
RS485 Gnd
Setpoint 1
Setpoint 2
Gnd
+
–
+
–
+
–
Total Reset
OC Total Pluse
Signal Gnd
RS485 B (–)
RS485 A (+)
Normally Open
Pushbutton
Switch
Figure 18: NEMA 4X hardware reset
Password
The password setting restricts access to the
Programming
and
Extended Programming
modes� Initially, the password is set to all zeros and any user can modify the parameter
settings� To change the password, press
ENTER
at the
Passwd
prompt� The first digit flashes�
Press
UP
to increment the digit and
RIGHT
to advance to the next digit� After entering all
digits, press
ENTER
to store the password and advance to
RstPswd
� The new password is
now required to enter either programming mode� With this password set, any user is able to
reset the stored totals on the monitor�