Page 18
Blancett B3100 Flow Monitor
DSY-UM-02266-EN-02_B3100 Automatic Unit Converter Tutorial.docx
ACTION
RESULT
N
OTICE
13. Press the PROG button
momentarily.
•
The PROG indicator
comes on.
Now you can set the decimal
pointer for the frequency only.
This decimals setting does not
change the MF setting.
14.
Press the ▲
button to
make the required
selection.
•
The selection 0.1 shows. For the explanation,
the decimals for the frequency (FR)
of the linearization points are set to
1 decimal.
15. Press the PROG button
momentarily.
•
The PROG indicator goes
off.
•
The selection is
confirmed.
-
3.2.2 Disable the linearization function
NOTICE
Best practice is to disable the linearization function if the function is not used.
ACTION
RESULT
N
OTICE
1. Press the PROG button
for at least 7 seconds.
•
The SETUP indicator
comes on continuously.
•
The setup menu shows.
-
2. Press the
►
button until
the LINEARIZE menu
shows.
•
The LINEARIZE menu
shows.
In this menu you can setup the
linearization or disable the
linearization function to save
battery power (if applicable).
3. Pres
s the ▲
button until
the LINEAR menu
shows.
•
The LINEAR menu
shows.
-
4. Press the PROG button
momentarily.
•
The PROG indicator
comes on.
Now you can enable or disable the
linearization function.
5.
Press the ▲
button to
make the required
selection.
•
The DISABLE selection
shows.
For the explanation,
DISABLE is selected.
6. Press the PROG button
momentarily.
•
The PROG indicator goes
off.
•
The selection is
confirmed.
-