Blancett B3100 Flow Monitor

 

DSY-UM-02266-EN-02_B3100 Automatic Unit Converter Tutorial.docx 

ACTION 

RESULT 

N

OTICE

 

12. 

Press the ▲

 button to 

make the required 

selection. 

  The selection US GAL 

shows. 

For the explanation,  

the measurement unit for the 

(accumulated) total is set to 

US GAL. 

13.  Press the PROG button 

momentarily. 

  The PROG indicator 

goes off. 

  The selection is 

confirmed. 

The Flowmeter (K-factor) unit m

is 

automatically converted to show the 

total in US Gallons. 

14. 

Press the ▲

 button until 

the DECIMALS menu 

shows. 

  The DECIMALS menu 

shows. 

15.  Press the PROG button 

momentarily. 

  The PROG indicator 

comes on. 

Now you can set the decimal 

pointer for display purposes only. 

16. 

Press the ▲

 button to 

make the required 

selection. 

  The selection 0.1 shows.  For the explanation,  

the decimals for the (accumulated) 

total are set to 1 decimal. 

17.  Press the PROG button 

momentarily. 

  The PROG indicator 

goes off. 

  The selection is 

confirmed. 

 

NOTICE 

The K-factor and the K-factor decimals for the (accumulated) total shows as 

AUTO because the automatic unit conversion controls these parameters. 

 

NOTICE 

If 7 digits is not sufficient to show the (accumulated) total, you can multiply the 

reading by x10, x100, etc. This multiplication factor is set in the next menu. 

 

ACTION 

RESULT 

N

OTICE

 

18.  Press the 

 button until 

the FACTOR X menu 

shows. 

  The FACTOR X menu 

shows. 

19.  Press the PROG button 

momentarily. 

  The PROG indicator 

comes on. 

Now you can set the multiplication 

factor for the (accumulated) total 

which require more than 7 digits. 

20.  Press the 

 button to 

make the required 

selection. 

  The selection 10 shows. 

For the explanation,  

the multiplication factor for the 

(accumulated) total.is set to 10x. 

21.  Press the PROG button 

momentarily. 

  The PROG indicator 

goes off. 

  The selection is 

confirmed. 

Summary of Contents for Blancett B3100 Series

Page 1: ...Blancett B3100 Flow Monitor Page 1 DSY UM 02266 EN 02_B3100 Automatic Unit Converter Tutorial docx B3100 Flow Monitor DSY UM 02266 EN 02 May 2021 TUTORIAL AUTOMATIC UNIT CONVERSION...

Page 2: ...g from this E112 series tutorial may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying recording or any information storage or retrieval system w...

Page 3: ...the flowmeter 5 2 1 1 The automatic unit conversion AUTO VOL and AUTO MAS mode 5 2 1 2 The automatic unit conversion HAND mode 7 2 2 Set the reading for the accumulated total 7 2 2 1 The accumulated t...

Page 4: ...r the accumulated total or the flowrate the automatic unit conversion uses the K Factor of the flowmeter to calculate the correct reading for the accumulated total and the flowrate separately NOTICE W...

Page 5: ...ed for 6624 605 average K factor m3 hr flow rate ACTION RESULT NOTICE 1 Press the PROG button for at least 7 seconds The SETUP indicator comes on continuously 2 Wait for the setup menu to show The set...

Page 6: ...the PROG button momentarily The PROG indicator comes on Now you can set the K Factor as given on the flowmeter certificate 14 Press the button and the button to make the required selection The select...

Page 7: ...Press the PROG button momentarily The PROG indicator goes off The selection is confirmed 8 Press the button to make the required selection The FLOWMETER menu shows 2 2 Set the reading for the accumula...

Page 8: ...als for the accumulated total are set to 1 decimal 17 Press the PROG button momentarily The PROG indicator goes off The selection is confirmed NOTICE The K factor and the K factor decimals for the acc...

Page 9: ...ent unit for the accumulated total A change of the measurement unit requires also a new calculation of the K Factor for the accumulated total The calculation is not done automatically 4 Press the butt...

Page 10: ...pointer 16 Press the button to make the required selection The selection 0 shows For the explanation the decimals for the K Factor are set to 0 decimals 17 Press the PROG button momentarily The PROG i...

Page 11: ...utton to make the required selection The selection US GAL shows For the explanation the measurement unit for the flow rate is set to US GAL 5 Press the PROG button momentarily The PROG indicator goes...

Page 12: ...s 1kHz use a setting above 50 pulses 16 Press the button to make the required selection The selection PLS 10 shows For the explanation the flow rate calculation is set to PLS 10 for 10 pulses per meas...

Page 13: ...atically 4 Press the button to make the required selection The selection US GAL shows For the explanation the measurement unit for the flow rate is set to US GAL 5 Press the PROG button momentarily Th...

Page 14: ...0 decimals 21 Press the PROG button momentarily The PROG indicator goes off The selection is confirmed 22 Press the button until the CALCULATE menu shows The CALCULATE menu shows The flow rate is calc...

Page 15: ...or comes on Now you can set the cut off time in seconds 28 Press the button and the button to make the required selection The selection 15 0 shows For the explanation the cut off time is set to 15 sec...

Page 16: ...log and the pulse output more accurate The linearization points defined by the frequency and meter factor are programmed into the B3100 to follow the flowmeter K Factor curve as close as possible The...

Page 17: ...he frequency 49 9 shows For the explanation the frequency of the linearization point is set to 49 9 Hz 6 When the frequency is entered press the button to make the required selection The PROG indicato...

Page 18: ...TICE Best practice is to disable the linearization function if the function is not used ACTION RESULT NOTICE 1 Press the PROG button for at least 7 seconds The SETUP indicator comes on continuously Th...

Page 19: ...Blancett B3100 Flow Monitor Page 19 DSY UM 02266 EN 02_B3100 Automatic Unit Converter Tutorial docx...

Page 20: ...Blancett B3100 Flow Monitor DSY UM 02266 EN 02_B3100 Automatic Unit Converter Tutorial docx...

Reviews: