Horizontal Mounting with Tubes Up
The mounting orientation shown in
is recommended for gas or slurry applications where condensate may
be present .
1
2
2
3
1
Flow
Flow
Figure 13: Tubes up (gases and slurries)
1
Supports—
Customer supplied
Rigid pipe supports approximately 3 and 6 pipe diameters from the end of the sensor
2
Isolation Valves—
Customer supplied
Full port ball
3
Ground
Protective (earth), 10 AWG (4 mm
2
) minimum
Horizontal Position with Flexible Armored Hoses
Flow
Flow
1
1
2
2
3
3
Figure 14: Flexible armored hoses
1
Supports—
Customer supplied
Rigid pipe supports approximately 3 and 6 pipe diameters from the end of the sensor
2
Hose—
Customer supplied
Flexible armored
3
Isolation Valves—
Customer supplied
Full port ball
Horizontal Position with Flexible Armored Hose Loops
Flow
Flow
1
1
3
3
2
2
Figure 15: Flexible armored hose loops
1
Supports—
Customer supplied
Rigid pipe supports approximately 3 and 6 pipe diameters from the end of the sensor
2
Hose—
Customer supplied
Flexible armored
3
Isolation Valves—
Customer supplied
Full port ball
Sensor Installation
Page 12
April 2019
CRL-QS-01552-EN-05