Vertical Mounting with Tubes to the Side, Flow Going Up
The mounting orientation shown in
is recommended for self-draining configurations .
1
1
3
2
2
Flow
Flow
Figure 16: Tubes to the side with flow going up (self-draining)
1
Supports—
Customer supplied
Rigid pipe supports approximately 3 and 6 pipe diameters from the end of the sensor
2
Isolation Valves—
Customer supplied
Full port ball
3
Ground
Protective (earth), 10 AWG (4 mm
2
) minimum
Sensor Installation
Page 13
April 2019