Sensor Mounting Positions and Locations
• Always mount the sensor downstream of the flow .
• Use vibration-reducing pipe supports approximately 3 and 6 pipe diameters from the end of the sensor in all applications .
• Install the sensor in a section of the piping where it always remains full, unless the application is designed to drain
with no flow . See
Vertical Mounting with Tubes to the Side, Flow Going Down
The mounting orientation shown in
is recommended for installation in an open vertical pipeline . If you MUST use
this configuration, make sure to use an isolation valve or other pipe restriction to prevent the sensor from running empty
while measurement is being taken .
Flow
Flow
1
2
3
4
5
Flow
1
Supply Tank
2
Sensor
3
Pipe Restriction
4
Valve
5
Batching Tank
Figure 10: Mounting positions and locations
Figure 11: Tubes to the side with flow going down)
Horizontal Mounting with Tubes Down
The mounting orientation shown in
is recommended for liquid applications .
1
1
2
2
3
Flow
Flow
Figure 12: Tubes down (liquids)
1
Supports—
Customer supplied
Rigid pipe supports approximately 3 and 6 pipe diameters from the end of the sensor
2
Isolation Valves—
Customer supplied
Full port ball
3
Ground
Protective (earth), 10 AWG (4 mm
2
) minimum
Sensor Installation