background image

Installation Instructions

To prepare for meter installation, follow these steps:
1   Close the meter's inlet-side valve  
2   Open a faucet and wait until water flow stops, to depressurize the system  Do not remove the meter until the  

flow stops  

3   Check valves and make necessary repairs to the curb (shutoff) valve or inlet side valve if necessary  
4   Before installing or removing a meter, close the outlet-side valve to relieve pressure  Protect the area around the meter 

against potential spills or leaks that could occur  

5   To replace an existing meter continue with Step 6  To install a new meter skip to Step 8 
6   Loosen the meter couplings or flange bolts and remove the meter and the old gaskets in the coupling nuts  You may 

choose to replace the old gaskets with the provided 9/64 in  thick rubber gaskets  

7   Clean the coupling nuts or flange bolts, removing any pipe dope or dirt from the threads or flange bolts  
8   Check the existing position for proper alignment and spacing  Correct any misalignment or spacing issues  
9   Place the connection gaskets inside the connection coupling nuts  
10  Install the meter in the pipeline in a horizontal or vertical position so that the flow arrow on the meter housing  

points in the same direction as water flow  Registration should be upright  

11.  For DN15…DN32 

a   Start the coupling nuts at the threaded meter ends  Verify that the nuts are properly aligned to avoid cross-threading 

or damage to the meter ends  This is especially important for the engineered polymer meter 

b   An effective method for starting a coupling nut is:

i   Position the nut squarely against the meter's spud end  
ii   Turn the nut counterclockwise (in reverse) while holding the nut against the meter spud end  When the first  

threads on both the nut and the spud end coincide, you will hear a slight click and feel the nut move into the  

starting position  

iii   Tighten the nut by hand until it is "hand-tight" 
iv   With an open-end wrench, apply a partial turn  Do not over tighten  For plastic swivel connections, a one-quarter 

turn beyond hand-tight is usually sufficient  

11. For  DN40 and DN50

a   With meter and gaskets in place, tighten the flange connection bolts  Verify the nuts are properly aligned to avoid 

damage to the flanged ends 

12  After the meter is installed, open the inlet shutoff valve until the meter is full of water and ensure that there are no leaks  

(The more flow you allow through the meter, appropriate for the meter size, the better ) 

13  Open the outlet valve until air is out of the meter and service line  
14  Open a valve downstream of the meter and verify that no foreign debris in the water obstructs the operations of  

the system  

15  Make sure the meter is installed with the

 flow arrow on the meter pointing in the direction of flow

  Check the read on the 

meter to make sure it is registering a positive number  If it is not, make sure the meter is installed in the correct direction  
a   The meter is sent in Storage mode so that customers do not experience alarms during shipment or installation  In 

general, a meter may take up to 2 minutes to begin measurement once the meter senses a full pipe  

b   The meter itself does not require a quantity of flow to begin measurement, the meter just requires that the pipe is 

cleared of air and filled with water  If the customer is attempting to purge the meter at low flow rates, it would likely be 

more difficult and take longer   

16  When the meter starts recording positive flow, note the meter read for your records 

E-Series® Ultrasonic Meters, Cold Water Meters

Page 8 

September 2017

ESM-UM-00255-EN-01

Summary of Contents for DN15

Page 1: ...E Series Ultrasonic Meters Cold Water Meters DN15 DN20 DN25 DN32 DN40 DN50 ESM UM 00255 EN 01 September 2017 User Manual...

Page 2: ...E Series Ultrasonic Meters Page ii September 2017 ESM UM 00255 EN 01...

Page 3: ...Meter 7 Special Fittings and Accessories 7 Installing E Series Ultrasonic Meters 7 Outdoor Installations 7 Indoor Installations 7 Installation Instructions 8 Protect Against Leakage 9 E Series Ultras...

Page 4: ...E Series Ultrasonic Meters Page iv September 2017 ESM UM 00255 EN 01...

Page 5: ...pment PRODUCT UNPACKING AND INSPECTION Upon opening the shipping container visually inspect the product and applicable accessories for any physical damage such as scratches loose or broken parts or an...

Page 6: ...al and reading Verify that a suitable electrical grounding wire is properly attached to the upstream and downstream pipe connections of the meter The grounding wire provides an alternative path for an...

Page 7: ...edded to ensure that it is not axially misaligned and that it lays evenly on the bottom of the pipe trench The backfill material covering the pipe should be placed appropriately to maintain pipe align...

Page 8: ...e nut against the meter spud end When the first threads on both the nut and the spud end coincide you will hear a slight click and feel the nut move into the starting position iii Tighten the nut by h...

Page 9: ...cape 7 Once water is flowing normally turn off the faucet E SERIES ULTRASONIC METER OPERATIONS Meter Display The Badger Meter E Series Ultrasonic meters use a nine digit Liquid Crystal Display LCD tha...

Page 10: ...feet or 100 thousand cubic meters with a resolution of 0 0001 cubic meters For DN40 and DN50 meters totalized flow displays up to 100 million gallons with a resolution of 0 1 gallons 10 million cubic...

Page 11: ...345678 10 gallons 123456 10 gallons 1234567 100 gallons 12345 100 gallons 123456 Typical Billing Units 1000 gallons 1234 Typical Billing Units 1000 gallons 12345 Cubic Feet Meter reading 1000th ft3 12...

Page 12: ...ding 7 digit reading Itron 100W 8 9 digit reading 6 digit reading 7 digit reading Itron 100W 9 digit reading 6 digit reading 7 digit reading No longer for sale 100W will transmit 9 digits through Itro...

Page 13: ...rm is sent to the endpoint NOTE No consumption data is sent to endpoint when the alarm is active Consumption data is sent to the endpoint except when Exceeding Max Flow alarm is set Reverse flow The m...

Page 14: ...o provide long term service with no maintenance The enclosure which includes the electronic meter s ultrasonic sensors battery and display is completely potted permanently sealed and non removable E S...

Page 15: ...m3 hr 4 0 gpm 1 71 m3 hr 7 5 gpm 1 71 m3 hr 7 5 gpm 1 71 m3 hr 7 5 gpm 3 2 m3 hr 14 0 gpm 4 5 m3 hr 20 gpm R Value 250 250 250 250 250 250 250 Operating Performance In the normal temperature range of...

Page 16: ...de C V Pedro Luis Ogaz n N 32 Esq Angelina N 24 Colonia Guadalupe Inn CP 01050 M xico DF M xico 52 55 5662 0882 Europe Eastern Europe Branch Office for Poland Latvia Lithuania Estonia Ukraine Belarus...

Reviews: