BATTERY REPLACEMENT
Battery powered monitors use a single 3 6V DC D size, lithium battery When replacement is necessary, use a clean fresh
battery to ensure continued trouble-free operation
Replacement Batteries
Manufacturer
Part Number
Badger Meter
B300028
Xeno
S11-0205-10-03
Tadiran
TL-5930/F
2 Unscrew the four captive screws on the front panel to gain access to the battery (see
)
Freq. In
4-20mA
Iso Total Pluse
TR_B
TR_A
RS485 Gnd
Setpoint 1
Setpoint 2
Gnd
+
–
+
–
+
–
Total Reset
OC Total Pluse
Signal Gnd
+ –
KB/Display
J1
TB1
JP1
JP2
JP3
P2
Input
Total P
ulse
Sig
nal
P1
Mag
Pulse
Iso
OC
Low
High
2
1
Unscrew
Captive Screws
Figure 17: Opening the
3 Press the tab on the battery connector to release it from the circuit board (see
)
Freq. In
4-20mA
Iso Total Pluse
TR_B
TR_A
RS485 Gnd
Setpoint 1
Setpoint 2
Gnd
+
–
+
–
+
–
Total Reset
OC Total Pluse
Signal Gnd
+ –
KB/Display
J1
TB1
JP1
JP2
JP3
P2
Input
Total P
ulse
Sig
nal
P1
Mag
Pulse
Iso
OC
Low
High
TR-A
TR-B
RS485 Gnd
Setpoint 1
Setpoint 2
Gnd
+
+
-
+
-
-
Total Reset
OC Total Pulse
Signal Gnd
Freq. In
4-20 mA
ISO Total Pulse
W
ARNING-
To r
educ
e the r
isk of ig
nition
of a flammable a
tmospher
e, ba
tt
er
ies
must only be changed in an ar
ea k
no
wn
to be nonflammable
.
REPLA
CE
BA
TTER WITH
XEON S11-0205-10-03
OR T
ADIR
AN TL
-5930/F
ONL
Y
Battery
Connector
Press
Release
Tab
Slide
Connector
Away
Figure 18: Battery connection release.
4 Remove the old battery and replace it with new one
5 Re-fasten the front panel screws
Battery Replacement
DSY-PM-00028-EN-02
Page 29
February 2017
Summary of Contents for ER-500
Page 1: ...Display ER 500 Advanced DSY PM 00028 EN 02 February 2017 Programming Manual...
Page 2: ...ER 500 Flow Monitor Page ii February 2017 DSY PM 00028 EN 02...
Page 4: ...ER 500 Flow Monitor Page iv February 2017 DSY PM 00028 EN 02...
Page 34: ...Display ER 500 Advanced INTENTIONAL BLANK PAGE DSY PM 00028 EN 02 Page 34 February 2017...
Page 35: ...Programming Manual INTENTIONAL BLANK PAGE DSY PM 00028 EN 02 Page 35 February 2017...