HR-LCD 4-20 scaled/unscaled

HR-LCD 4-20 SCALED/UNSCALED

The HR-LCD 4-20 scaled/unscaled is a permanently sealed, electronic LCD register that produces a scaled/unscaled output as 

well as an analog 4-20 mA DC output signal with a dual output wire design .

 

Refer t

Figure 2

 .

Wire Connections

•  Scaled/unscaled side cable (

Flying lead for field splice connection)

◊  For a scaled output, connect the 

red

 wire (positive) and the 

black

 wire (negative) .

◊  For an unscaled output, connect the 

green

 wire (positive) and the 

black

 wire (negative) .

OTE: 

N

The unscaled output will be active only when the device is powered by the 4-20 output .

•  4-20 mA side cable (

Flying lead for field splice connection)

Scaled/Unscaled Cable

4-20 mA Cable

RED - Scaled Pulse

GREEN - Unscaled Pulse

BLACK - Ground

RED - Loop/External Power + 

BLACK - Loop Return/External Power 

Figure 2:  HR-LCD 4-20 scaled/unscaled wiring

Measurement Resolution

Recordall 

 Disc Series

Size (in.)

Scaled  

(pulse/unit)

Unscaled (pulse/unit)

Analog Output

gal

ft

3

m

3

gal

ft

3

m

3

20 mA Set point 

(gpm)

LP

5/8

1

10

100

228 .415

1708 .661

60337 .105

20

M25

5/8

1

10

100

198 .334

1483 .641

52391 .084

25

M35

3/4

1

10

100

126 .678

947 .621

33462 .863

35

M40

1

1

10

100

89 .783

671 .621

23716 .632

40

M55

1

1

10

100

58 .065

434 .358

15338 .279

55

M70

1

1

10

100

46 .773

349 .884

12355 .278

70

M120

1-1/2

0 .10

1

10

23 .866

178 .533

6304 .435

120

M170

2

0 .10

1

10

14 .565

108 .950

3847 .303

170

Recordall  

Turbo Series

Size (in.)

Scaled  

(pulse/unit)

Unscaled (pulse/unit)

Analog Output

gal

ft

3

m

3

gal

ft

3

m

3

20 mA Set point 

(gpm)

T160

1-1/2

0 .10

1

10

1 .537

11 .494

405 .894

200

T200

2

0 .10

1

10

1 .537

11 .494

405 .894

310

T450

0 .10

1

10

1 .598

11 .954

422 .109

550

T1000

4

0 .10

1

10

1 .665

12 .455

439 .820

1250

T2000

6

0 .01

0 .10

1

0 .150

1 .123

39 .639

2500

T3500

8

0 .01

0 .10

1

0 .151

1 .131

39 .939

4500

T5500

10

0 .01

0 .10

1

0 .198

1 .481

52 .308

7000

T6200

12

0 .001

0 .01

0 .10

0 .129

0 .963

34 .006

8800

T6600

16

0 .001

0 .01

0 .10

0 .016

0 .116

4 .107

13200

T1000

20

0 .001

0 .01

0 .10

0 .009

0 .067

2 .382

19800

Recordall 

Compound 

Series

Size (in.)

Scaled (pulse/unit)

Unscaled (pulse/unit)

Analog Output

gal

ft

3

m

3

gal

ft

3

m

3

20 mA Set point 

(gpm)

High Side T200

2

0 .10

1

10

1 .537

11 .494

405 .894

200

Low Side M25

2

1

10

100

198 .334

1483 .641

52391 .084

25

High Side T450

3

0 .10

1

10

1 .598

11 .954

422 .109

450

Low Side M25

3

1

10

100

198 .334

1483 .641

52391 .084

25

High Side T1000

4

0 .10

1

10

1 .665

12 .455

439 .820

1000

Low side M35

4

1

10

100

126 .678

947 .621

33462 .863

35

High Side T2000

6

0 .01

0 .10

1

0 .150

1 .123

39 .639

2000

Low Side M35

6

1

10

100

126 .678

947 .621

33462 .863

35

High Side T3500

8

0 .01

0 .10

1

0 .151

1 .131

39 .939

Low side M120

8

0 .10

1

10

23 .866

178 .533

6304 .435

Page 8 

April 2020

REG-UM-02466-EN-02

Summary of Contents for HR-LCD 4-20

Page 1: ...High Resolution LCD Registers HR LCD 4 20 scaled unscaled HR LCD Pulse REG UM 02466 EN 02 April 2020 User Manual...

Page 2: ...High Resolution LCD Registers HR LCD 4 20 scaled unscaled HR LCD Pulse Page ii April 2020 REG UM 02466 EN 02...

Page 3: ...re 6 9 Digit Totalization 6 Rate of Flow 7 Meter Model Information 7 Installing the Register 7 Bayonet Mount 7 HR LCD 4 20 scaled unscaled 8 Wire Connections 8 Measurement Resolution 8 Specifications...

Page 4: ...High Resolution LCD Registers HR LCD 4 20 scaled unscaled HR LCD Pulse Page iv April 2020 REG UM 02466 EN 02...

Page 5: ...ity to operate the equipment DESCRIPTION High resolution registers are fully electronic solid state devices with no moving parts The devices come standard as factory programmed with the option for pro...

Page 6: ...oot or cubic meter Example 123456 displayed value x 10 multiplier value 1234560 Extended Visual Reading Resolution Typical Billing Segments Status Indicators IR Programming Port 00 00 4 Digit Date Cod...

Page 7: ...appears and Compound meters are indicated by a C See examples below Model 25 Disc Series Meter Calibrated in Gallons Model 450 Turbo Series Meter Calibrated in Gallons 2 in Low Side Compound Series M...

Page 8: ...38 279 55 M70 1 1 10 100 46 773 349 884 12355 278 70 M120 1 1 2 0 10 1 10 23 866 178 533 6304 435 120 M170 2 0 10 1 10 14 565 108 950 3847 303 170 Recordall Turbo Series Size in Scaled pulse unit Unsc...

Page 9: ...black wire negative Scaled digital output from the register has a default resolution of 1 10th of the register test circle resolution may vary in some cases The movement of the meter magnet is conver...

Page 10: ...ies Size in Scaled pulse unit gal ft3 m3 LP 5 8 1 10 100 M25 5 8 1 10 100 M35 3 4 1 10 100 M40 1 1 10 100 M55 1 1 10 100 M70 1 1 10 100 M120 1 1 2 0 10 1 10 M170 2 0 10 1 10 Recordall Turbo Series Siz...

Page 11: ...nt 100 mA Pulse Width 50 ms programmable 30 100 ms Power Input Voltage Range 9 50V DC supply Max Load Resistance Ohms 50 Ohms 50 Ohms supply voltage 9V Battery Lithium thionyl chloride AA cell fully e...

Page 12: ...alarm remains active for 35 days The alarm automatically clears after 35 days if reverse flow condition has not recurred Suspected leak Register detects 24 hours without one 15 minute interval of no...