Flow meter ModBus
®
register table 2017-06-29, Version 2.00
Page 36/43
UF_iS4000_BA_02_1811
0x0115 FLOAT
Read
only
MEASURE
Dry calibration
0x0119 U16
Read
only
MEASURE_COUNTER
Dry calibration
0x0125 U16
Admin
COMMAND
1: save configuration
2: restore configuration
6: save totalizers
7: clear totalizers
8: clear totalizers
14: current loop calibration point A
15: current loop calibration point B
16: current loop calibration complete
22: default save
23: remote reset
24: default restore
26: make file system
34: press key up
35: press key right
36: press key save exit
38: print screen
41: open channel –
calculate upper range
42: open channel – use default water
level
0x0126 FLOAT
Factory CURRENTLOOP_POINTA Dry calibration
0x0128 FLOAT
Factory CURRENTLOOP_POINTB Dry calibration
0x012A U16
Service SIMULATION
Not stored in non-volatile memory
0: 0.0%
10: + 10.0%
20: + 20.0%
30: + 30.0%
40: + 40.0%
50: + 50.0%
60: + 60.0%
70: + 70.0%
80: + 80.0%
90: + 90.0%
100: +100.0%
65408: Off
65436: -100.0%
65446: - 90.0%
65456: - 80.0%
65466: - 70.0%
65476: - 60.0%
65486: - 50.0%
65496: - 40.0%
65506: - 30.0%
65516: - 20.0%
65526: - 10.0%
0x012B U32
Read
only
RANDOM
Security
0x012E U16
Service
ALARM_MODE_OF_
ANALOG_OUTPUT
0: none
3: 23 mA
4: 3.5 mA