Page 52
4-11
Model M-2000 Mag Meter
Menu Manager Configuration Errors
Error
Description
Recommended Action
110
Output 1/2:
Pulse Output
Configuration Error
This error is observed when improperly configuring the full scale flow, pulse per unit, or pulse width.
This error can indicate the following configuration violations:
1.
Pulse frequency exceeds limits at full scale flow
2.
Pulse duty cycle is less than 50% at full scale flow (pulse on time > pulse off time)
Pulse frequency limit is 10 kHz. However, with a non-zero pulse width configuration, the limit is 500
Hz to achieve a 50% pulse duty cycle.
If not using the pulse outputs, set the pulses per unit to zero to allow for reconfiguration of the full
scale flow. If it is required to use the pulse outputs, reevaluate the pulse output configuration.
Consider recording and clearing totalizers prior to changing totalizer units.
120
Display:
Totalizer Conversion
Error – Totalizer cannot be properly
converted for display
This error is observed while trying to change the totalizer units. Limits of display will prevent
improper configuration of the volume unit dependent on current totalizer values. Consider
recording and cleaning totalizers prior to changing totalizer.
121
Output 1/2:
Pulse Output
Configuration Error
This error is observed when changing the totalizer units of measure. This error implies the pulse
configuration exceeds limits (see error 110). Please note the pulses per unit is not automatically
updated on volume unit reconfiguration. The pulses per unit should be manually changed to
accommodate the desired units of measure. It may be necessary to set the pulses per unit to zero
then change to the desired totalizer units.
140
Output 3:
Configuration Error – Full
scale frequency exceeds limits of
relay (1000 Hz)
Reduce full scale frequency setting of output when hardware is configured for relay operation.
150
Output 3:
Configuration Error – Full
scale frequency exceeds limits (10
kHz)
Reduce full scale frequency setting of output when hardware is configured for open collector
operation.
Display Error / Status Messages
Error Message
Possible Cause
Recommended Action
Err: Detector
No detector connection with amplifier.
Check detector and cable connections in accordance with this manual.
Connection between amplifier and detector
Contact Technical Support.
Supply voltage too low.
Contact Technical Support.
Grounded coils in meter.
Contact Technical Support.
Water in detector.
Contact Technical Support.
Err: Empty pipe
Pipe may not be full.
Make sure all trapped air is out of system.
If fluid or fluid conductivity recalibrate the parameter.
Err: Full scale
Actual flow rate is exceeding programmed flow. Reduce flow rate or increase the programmed full scale value by more than 5%.
Err: AD-Range
AD-Converter is exceeding signal limits.
Check the grounding scheme of the meter installation. See grounding section in
this manual. Verify pipe is not empty.
Err: AD-INT
Initialization of AD-Converter unsuccessful.
Contact Technical Support.
Err: Rollover
Rollover counters have exceeded limit.
Clear all totalizers.
Err: Rollover Status
Totalizer rollover has occurred.
Reload totalizer then clear all totalizers.
Err: Simulation
I/O simulator is enabled.
Disable simulator in I/O menu.