Programming
Page 25/47
MID_M1000_BA_02_1802
Digital Outputs
Functional
selection
The following functions can be selected for the
Outputs 1 to 2 as well as for the Solid State Relay.
The Solid State Relay function is linked with the func-
tion of output 2.
Off
means digital output is switched off.
Forward pulse
generates pulses during forward flow
conditions.
Reverse pulse
generates pulses during reverse flow
conditions.
Min/Max Alarm
provides indication when flow rate
exceeds thresholds defined by Set Min. or Set Max. in
% of full scale.
Empty pipe alarm
provides indication when pipe is
empty.
Flow direction
provides indication on current flow
direction
Preset
provides indication when preset batch amount
has been realized.
Frequency
generates a defined frequency of full
scale.
Error alarm
provides indication when meter has error
condition.
Loopback
shows the status of the digital input
Test
is only used for the Verification Device
Function
Out1 Out2 / Solid
State Relay
Off
X
X
Forward pulse
X
X
Reverse pulse
X
X
Min/Max Alarm
X
X
Empty pipe
X
X
Flow direction
X
X
Preset
X
X
Error alarm
X
X
Frequency
X
X
Loopback
X
X
Test
X
X